DOI 10.1007/s00062-026-01676-y
In-Stent Stenosis After Surface-Modified Versus Uncoated Flow Diversion: a Randomized Trial (Derivo 2 Heal vs Derivo 2).
Аннотация
BACKGROUND AND PURPOSE: Surface-modified flow diverters have been introduced to reduce thrombogenicity and potentially influence vascular healing. However, their impact on in-stent stenosis and aneurysm occlusion remains uncertain. We compared a coated flow diverter (Derivo 2 heal) with the corresponding uncoated device (Derivo 2) regarding in-stent stenosis development and aneurysm occlusion in a randomized clinical setting.
MATERIALS AND
METHODS
In this single-center randomized, open-label trial, 21 patients with unruptured internal carotid artery aneurysms were treated with either a coated (n = 10) or uncoated (n = 11) flow diverter. All patients received dual antiplatelet therapy and underwent digital subtraction angiography at 3 and 9 months. In-stent stenosis was assessed using a standardized ratio-based method. Device characteristics and oversizing were recorded. The presumed etiology of in-stent stenosis was categorized as intimal hyperplasia or stent deformity based on morphologic appearance.
RESULTS
Mean in-stent stenosis at 3 and 9 months did not differ between groups. Complete aneurysm occlusion at 9 months was lower in the coated group (40.0% vs 81.8%; two-sided Fisher's exact p = 0.080). Oversizing was not associated with in-stent stenosis but was numerically greater in patients with complete occlusion. Stent deformity was associated with higher in-stent stenosis than presumed intimal hyperplasia (3 months: 40.8% vs 21.7%, p = 0.007; 9 months: 30.1% vs 13.3%, p = 0.022).
CONCLUSIONS
In this randomized single-center study, the surface-modified Derivo 2 heal did not demonstrate reduced in-stent stenosis compared with the uncoated Derivo 2, while complete occlusion at 9 months was lower in the coated-device group. These findings warrant confirmation in larger comparative studies.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.