Embolization Followed by Self-Repositioning of a CardioMEMS™ Wireless Pulmonary Artery Hemodynamic Monitoring Sensor Device.

Kharsa Chloe, Qamar Fatima, Elias Joseph, Abou Zeid Karl, Goel Sachin S, Aoun Joe

Клинический случайСтатьяПеревод в очереди

Аннотация

We report a case of early migration of a CardioMEMS™ heart failure device (Abbott) from the right to the left pulmonary artery, which then spontaneously repositioned itself within the left pulmonary artery without the need for intervention. A balloon Swan-Ganz catheter was then used to reposition the device, successfully lodging it into a more distal position. This case underscores the importance of close device monitoring and demonstrates an effective technique for repositioning without device retrieval.

Для цитирования: Kharsa Chloe, Qamar Fatima, Elias Joseph, Abou Zeid Karl, Goel Sachin S, Aoun Joe. Embolization Followed by Self-Repositioning of a CardioMEMS™ Wireless Pulmonary Artery Hemodynamic Monitoring Sensor Device.. Methodist DeBakey cardiovascular journal. 2026. DOI: 10.14797/mdcvj.1765.