First use of an expandable sheath and transcaval access for transcatheter Edwards SAPIEN 3 aortic valve implantation.

Ott Ilka, Greenbaum Adam B, Schunkert Heribert, Kastrati Adnan, Kasel Markus

Клинический случайСтатьяVideo-Audio MediaПеревод в очереди

Аннотация

AIMS: We aimed to describe the first percutaneous transcaval TAVI in Europe using the expandable introducer sheath (eSheath; Edwards Lifesciences, Irvine, CA, USA) for implantation of an Edwards SAPIEN 3 aortic valve (23 mm).

METHODS AND

RESULTS

A 72-year-old male patient presented with dyspnoea (NYHA III) in part due to severe aortic stenosis. Concomitant diseases were severe peripheral artery disease, severe chronic obstructive pulmonary disease (GOLD IV) and renal insufficiency, such that neither a transfemoral nor a transapical approach was considered favourable. An eSheath was introduced into the abdominal aorta via the femoral vein and the inferior vena cava. TAVI was then performed according to standard procedures. After TAVI, the veno-arterial junction was occluded with a 6 mm AMPLATZER Muscular VSD Occluder (St. Jude Medical, St. Paul, MN, USA) using an 8.5 Fr Agilis™ sheath (St. Jude Medical) inside the TAVI sheath. Aortography showed no retroperitoneal accumulation of contrast media. The femoral vein access site was closed using two prepositioned sutures (ProGlide; Abbott Vascular, Santa Clara, CA, USA). A CT scan the following day showed no retroperitoneal haemorrhage.

CONCLUSIONS

Percutaneous transcaval venous access to the aorta may provide a new access strategy for TAVI in otherwise ineligible patients and appears safe with expandable sheath technology.

Для цитирования: Ott Ilka, Greenbaum Adam B, Schunkert Heribert, Kastrati Adnan, Kasel Markus. First use of an expandable sheath and transcaval access for transcatheter Edwards SAPIEN 3 aortic valve implantation.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2015. DOI: 10.4244/eijy15m05_08.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.