DOI 10.1002/ccd.31448

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2025

Optimizing Device Selection in Percutaneous Paravalvular Leak Closure: A Comparative Study of Different Transthoracic and Transesophageal Echocardiographic Techniques.

Tekin Meltem, Güler Gamze Babur, Çiçek Mehmet, Tanboğa İbrahim Halil, Pysz Piotr, Güler Arda, Demir Ali Rıza, Efe Yusuf, Atmaca Sezgin, Pay Dilara, Kalkan Ali Kemal, Ertürk Mehmet

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Clinically significant paravalvular leak (PVL) may complicate both surgical and transcatheter valve implantation. Percutaneous PVL closure (PPVLC) is becoming an increasingly attractive alternative to redo surgery, with demonstrated lower mortality and morbidity rates. Echocardiographic techniques are crucial for accurate diagnosis, defect sizing, and determining the appropriate size of the sealing devices.

AIMS: There is no consensus on the optimal imaging modality for PVLs. We aimed to compare transthoracic and transesophageal echocardiographic measurements to accurately determine defect size and estimate device size.

METHODS

We reviewed hospital records to identify patients diagnosed with moderate to severe and severe PVL from 2018 to 2024. A total of 81 patients who underwent PPVLC were evaluated. Eight of these patients were excluded due to unsuccessful PPVLC, leaving 73 patients who were successfully treated. The defect size for all patients was evaluated using 2D transthoracic echocardiography (TTE), 2D transesophageal echocardiography (TEE), direct 3D TEE cropped volume rendering vena contracta (VC) measurement, and 3D TEE multiplanar reconstruction (3D TEE MPR).

RESULTS

Among the 73 patients, 42 underwent aortic PPVLC and 31 underwent mitral PPVLC. Proportional odds logistic regression analysis identified 3D TEE MPR measurement as the strongest predictor of device size accuracy, both overall and within the aortic/mitral subgroups. In the mitral subgroup, the predictive power of direct 3D TEE cropped volume rendering VC measurement and 3D TEE MPR measurement were similar. Furthermore, a cut-off value of 7 mm was identified for hemodynamically significant jets as measured by 3D methods.

CONCLUSION

Our findings suggest that using 3D TEE MPR significantly improves the accuracy of device size selection in both mitral and aortic PVL. Additionally, direct 3D TEE cropped volume rendering VC measurement can serve as a viable alternative for patients with mitral PVL.

Для цитирования: Tekin Meltem, Güler Gamze Babur, Çiçek Mehmet, Tanboğa İbrahim Halil, Pysz Piotr, Güler Arda, Demir Ali Rıza, Efe Yusuf, Atmaca Sezgin, Pay Dilara, Kalkan Ali Kemal, Ertürk Mehmet. Optimizing Device Selection in Percutaneous Paravalvular Leak Closure: A Comparative Study of Different Transthoracic and Transesophageal Echocardiographic Techniques.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.31448.