DOI 10.1097/ijg.0000000000000745

Медицинская библиотека · Journal of glaucoma · 2017

Comparison of Central Corneal Thickness Measurements Using Ultrasonic Pachymetry, Anterior Segment OCT and Noncontact Specular Microscopy.

Scotto Riccardo, Bagnis Alessandro, Papadia Marina, Cutolo Carlo Alberto, Risso Domenico, Traverso Carlo Enrico

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

PURPOSE: To evaluate and compare central corneal thickness (CCT) values measured with anterior segment optical coherence tomography (AS-OCT), noncontact specular microscopy (NCSM), and ultrasound pachymetry (USP).

MATERIALS AND

METHODS

CCT was measured in 182 healthy eyes without ocular abnormalities other than refractive errors. Three consecutive measurements of CCT by the same examiner were obtained during the same session. The testing sequence of AS-OCT and NCSM was randomly selected. The USP always was performed after the noncontact examinations.

RESULTS

The average CCT measured by AS-OCT, NCSM, and USP were 535.8±35.5, 547.7±38.2, and 537.4±37.5 μm, respectively. The mean differences between modalities were 11.8±14.7 μm (P<0.01) between NCSM and AS-OCT, 10.3±17.7 μm (P<0.01) between NCSM and USP and 1.6±8.6 μm (P=0.02) between USP and AS-OCT.

CONCLUSIONS

AS-OCT, NCSM, and USP showed an overall strong agreement in measuring CCT. However, CCT measurements with AS-OCT showed a good correlation to those obtained by USP, NCSM tended to give statistically significant higher CCT readings than either alternative and showed the worse repeatability indices. On the basis of our results, CCT measurement obtained with different instruments cannot be considered directly interchangeable.

Для цитирования: Scotto Riccardo, Bagnis Alessandro, Papadia Marina, Cutolo Carlo Alberto, Risso Domenico, Traverso Carlo Enrico. Comparison of Central Corneal Thickness Measurements Using Ultrasonic Pachymetry, Anterior Segment OCT and Noncontact Specular Microscopy.. Journal of glaucoma. 2017. DOI: 10.1097/ijg.0000000000000745.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.