DOI 10.1177/19433654261472943

Медицинская библиотека · Respiratory care · 2026

In Vitro Comparison of a Low-Cost Bubble-NIV and Ventilator-Driven NIV.

Pierce Pucci Josephine U, Pierce Bradley T, Pierce Lisa M, Shriver Nicholas L, Grete Alicyn F, Burgess Alex M, Minner Emily S, DiAngelis Dante M, Nowicki Margaret A, Ayer David, Barnhill Jason C, Naun Christopher A

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Almost 2.5 million neonatal deaths occur annually, disproportionately affecting low- and middle-income countries. Respiratory support devices remain largely inaccessible in these regions because of prohibitive costs and infrastructure requirements. The WHO-style bubble CPAP provides accessible continuous positive pressure but cannot deliver bi-level support when CPAP alone is insufficient. We hypothesized that a low-cost bubble noninvasive ventilatory (Bubble-NIV) system, meant to augment the WHO-style bubble CPAP, would perform non-inferiorly to the ventilator-driven RAM-NIV in delivering positive pressure across variable nasal interface conditions in bench testing.

METHODS

We developed a low-cost Bubble-NIV system (∼$50) with a servo motor-driven lever to drive bi-level pressures via variable distal tubing water submersion depths. Performance was compared with ventilator-driven RAM-NIV in a static in vitro model across 3 nasal cannula fits (80%, 70%, 60% nostril occlusion). Primary outcome was mean airway pressure measured in the nasopharynx.

RESULTS

Bubble-NIV maintained bi-level pressures and delivered pressures closer to target (mean airway pressure, mean goal: 15 cm H2O) than ventilator-driven RAM-NIV across all nasal interface scenarios: 15.1 versus 13.4 cm H2O at 80% occlusion (P < .001), 12.6 versus 7.4 cm H2O at 70% occlusion (P < .001), and 5.6 versus 1.8 cm H2O at 60% occlusion (P < .001). The performance advantage increased with worsening leak (1.1-fold to 3.1-fold). Optimal Bubble-NIV pressures occurred with flows set within 2 L/min of minimal bubbling. Oscillatory waveforms of Bubble-NIV were greater than RAM-NIV.

CONCLUSIONS

A low cost Bubble-NIV system demonstrated superior nasal leak tolerance compared with ventilator-driven RAM-NIV in bench testing. The low-cost, nonproprietary design addresses accessibility barriers to delivering bi-level noninvasive respiratory support for neonates in resource-limited settings.

Для цитирования: Pierce Pucci Josephine U, Pierce Bradley T, Pierce Lisa M, Shriver Nicholas L, Grete Alicyn F, Burgess Alex M, Minner Emily S, DiAngelis Dante M, Nowicki Margaret A, Ayer David, Barnhill Jason C, Naun Christopher A. In Vitro Comparison of a Low-Cost Bubble-NIV and Ventilator-Driven NIV.. Respiratory care. 2026. DOI: 10.1177/19433654261472943.