Agreement of Corvis ST Biomechanical IOP Measurements With Pachymetry-Based Correction Formulas in Patients With Keratoconus: A Comparative Analysis According to Disease Severity.

Aydın Fahri Onur, Ceylan Ali, Akbaş Yusuf Berk, Çelen Arabacı İdil, Ermiş Serhat, Kepez Yıldız Burçin

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

To evaluate the agreement of intraocular pressure (IOP) measurements obtained by Corvis ST using biomechanically corrected IOP (bIOP), noncontact tonometry (NCT), and pachymetry-based correction formulas (Ehlers, Shah, Dresden, Spoerl) across different keratoconus severity levels and in comparison with healthy control eyes.

METHODS

This prospective, cross-sectional study included 77 patients with keratoconus (77 eyes), categorized by severity using Pentacam-based Topographic Keratoconus Classification and 54 healthy control patients. Patients underwent corneal tomography, and IOP was measured with Corvis ST. Uncorrected NCT-IOP (IOPnct), biomechanically corrected IOP (bIOP), and pachymetry-based corrected IOP (Ehlers, Shah, Dresden, Spoerl) were compared. Agreement among methods was assessed using Bland-Altman analysis.

RESULTS

The mean patient age was 26.4±7.3 years (59.7% male). NCT consistently underestimated IOP compared with bIOP, while Ehlers (+2.6 mm Hg), Shah (+1.4 mm Hg), and Dresden (+0.7 mm Hg) formulas significantly overestimated ( P <0.001). The Spoerl formula (-0.9 mm Hg) showed no significant difference compared with bIOP ( P =0.209). Bland-Altman analysis confirmed that the Spoerl formula had the smallest bias and narrowest limits of agreement across all keratoconus stages. Differences between methods became more pronounced with increasing severity, particularly evident for Ehlers and Shah formulas. In healthy control eyes, the different IOP estimation methods showed relatively small bias relative to bIOP.

CONCLUSIONS

Noncontact tonometry significantly underestimates, and common pachymetry-based corrections (Ehlers, Shah, Dresden) tend to overestimate IOP in keratoconic eyes. The Spoerl formula provides IOP values most closely aligned with the Corvis ST bIOP across all keratoconus stages and was the most consistent pachymetric correction method relative to this biomechanically corrected comparator. Agreement between methods was closer in healthy corneas than in keratoconic eyes, supporting the influence of altered corneal biomechanics on tonometric measurements.

Для цитирования: Aydın Fahri Onur, Ceylan Ali, Akbaş Yusuf Berk, Çelen Arabacı İdil, Ermiş Serhat, Kepez Yıldız Burçin. Agreement of Corvis ST Biomechanical IOP Measurements With Pachymetry-Based Correction Formulas in Patients With Keratoconus: A Comparative Analysis According to Disease Severity.. Eye & contact lens. 2026. DOI: 10.1097/icl.0000000000001284.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.