Diagnostic value of Sirius Scheimpflug-Placido corneal tomography in keratoconus and subclinical keratoconus.

Dal Ali, Postacı Sevinç Arzu, Akikol Talip, Kutluksaman Bünyamin

СтатьяПеревод в очереди

Аннотация

PURPOSE: To determine the diagnostic accuracy and optimal threshold values of anterior and posterior corneal parameters obtained with the Sirius+ Scheimpflug-Placido tomography system in detecting clinical keratoconus (KCN) and subclinical keratoconus (sKCN).

METHODS

In this cross-sectional study, 112 eyes (40 KCN, 32 sKCN, and 40 healthy controls) were classified based on clinical and tomographic findings. Corneal tomography was performed using the Sirius+ system under standardized conditions. The evaluated parameters included anterior and posterior symmetry indices (SIf, SIb), anterior and posterior keratoconus vertex values (KVf, KVb), and minimum corneal thickness (THKmin). Diagnostic performance was assessed using receiver operating characteristic (ROC) curve analysis, including area under the curve (AUC), sensitivity, specificity, and optimal cut-off values.

RESULTS

THKmin progressively decreased from controls to sKCN and KCN (528.08 ± 36.26, 479.30 ± 32.55, and 455.75 ± 37.76 µm, respectively; p ≤ 0.001). Conversely, SIf and SIb values increased significantly across groups (SIf: 0.21±0.36, 1.19±0.82, 3.86±3.09 D; SIb: 0.05±0.07, 0.37±0.39, 1.01±0.65 D; all p ≤ 0.001). Similarly, KVf and KVb values showed a progressive increase, with KVb demonstrating marked elevation in KCN (p < 0.001). For differentiating sKCN from controls, SIb showed the highest diagnostic performance (AUC = 0.934), with 90.0% sensitivity and 85.0% specificity at a cutoff value of >0.11 D

Conclusion

Posterior corneal parameters derived from Sirius+ tomography, particularly SIb and KVb, demonstrated superior diagnostic performance in detecting both clinical and subclinical keratoconus. Posterior surface-based multiparametric analysis may improve early detection and should be incorporated into routine tomographic evaluation.

Для цитирования: Dal Ali, Postacı Sevinç Arzu, Akikol Talip, Kutluksaman Bünyamin. Diagnostic value of Sirius Scheimpflug-Placido corneal tomography in keratoconus and subclinical keratoconus.. Photodiagnosis and photodynamic therapy. 2026. DOI: 10.1016/j.pdpdt.2026.105508.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.