MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.pdpdt.2026.105395

Медицинская библиотека · Photodiagnosis and photodynamic therapy · 2026

Impact of multifocal contact lens wear on optical coherence tomography measurements of the cornea and retina.

Demirtaş Atılım Armağan, Turun Bozkurt Başak, Yüce Berna, Küsbeci Tuncay

СтатьяПеревод в очереди

Аннотация

PURPOSE: This study aimed to evaluate the influence of multifocal contact lenses (MFCLs) on structural optical coherence tomography (OCT) parameters, including central corneal thickness (CCT), ganglion cell-inner plexiform layer (GCIPL) thickness, and peripapillary retinal nerve fiber layer (RNFL) thickness. In addition, we assessed their effect on scan acquisition quality, as measured by the OCT quality index (QI).

METHODS

A retrospective cross-sectional study was performed on 80 eyes of 40 female presbyopic participants (mean age 52.4 ± 6.1 years). All subjects were fitted with Acuvue Oasys Multifocal lenses (Johnson & Johnson Vision). OCT imaging was performed for the anterior segment, ganglion, and disc analyses both with and without MFCLs. Parameters assessed included CCT, average and minimum GCIPL thickness, average RNFL thickness, and scan QI. Only scans with QI ≥ 4 were included. Correlation analyses were performed between OCT parameters and clinical parameters (age, spherical equivalent, K1, K2, and corneal radius), and MFCL variables (spherical power and addition levels).

RESULTS

Linear mixed model analysis revealed significant differences in several OCT parameters between lens-off and lens-on conditions. CCT slightly increased with MFCL wear (adjusted means of difference = +1.9 µm, p < 0.001), while anterior scan quality remained stable. In contrast, posterior segment scan quality decreased significantly in both ganglion (p = 0.003) and disc (p = 0.002) modules, accompanied by a reduction in average RNFL thickness (adjusted means of difference = -3.19 µm, p < 0.001). Correlation analyses showed that both spherical equivalent and MFCL spherical power were significantly associated with GCIPL thickness, suggesting a subtle optical influence of refractive status and lens power on retinal measurements. Addition levels showed no significant correlations with any OCT-derived parameters.

CONCLUSIONS

MFCL wear was associated with reduced OCT image quality in posterior segment imaging and with thinner RNFL measurements, while anterior segment quality and GCIPL parameters remained stable. Correlation analyses revealed that refractive status and lens spherical power subtly influenced GCIPL thickness. Therefore, OCT examinations in presbyopic patients should be interpreted with caution, and baseline scans without lens wear may be preferable, particularly in glaucoma screening.

Для цитирования: Demirtaş Atılım Armağan, Turun Bozkurt Başak, Yüce Berna, Küsbeci Tuncay. Impact of multifocal contact lens wear on optical coherence tomography measurements of the cornea and retina.. Photodiagnosis and photodynamic therapy. 2026. DOI: 10.1016/j.pdpdt.2026.105395.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.