MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3928/1081597x-20251217-02

Медицинская библиотека · Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026

Surgical Approach for the Implantation of a Ring Electrode into the Sulcus in a Phakic Non-human Primate.

Kortuem Friederike C, Seitz Immanuel, Kaltenstadler Sebastian, Schumayer Sven, Sigdel Bishesh, Peters Tobias, Rothermel Albrecht, Bucher Volker, Zrenner Eberhart, Bartz-Schmidt Karl-Ulrich, Straßer Torsten

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the feasibility of surgical implantation of electrical components needed for the operation of biomimetic artificial intraocular lenses (IOLs) for presbyopia correction in phakic non-human primates. This included positioning a ring electrode into the posterior chamber sulcus, securing a circuit board-mounted implant under the superior rectus muscle, and placing a battery pack in the orbit. The full implant allows for recording and transmission of electrical signals within the ciliary muscle during accommodation.

METHODS

The implant consists of a microcontroller, a biopotential amplifier, and a Bluetooth transmitter on a flexible printed circuit board (PCB), a battery, and the ring electrode. After mobilization and placement of traction sutures under three muscles, the PCB was positioned under the superior rectus muscle. The ring-shaped electrode was pushed into the anterior chamber via a clear cornea tunnel incision in the upper temporal quadrant. The electrode was carefully advanced under the iris with two forceps to prevent tilting and damaging the anterior capsule until its final position under the iris above the ciliary muscle.

RESULTS

During the implantation tests, in two eyes the ring electrode broke during the advancement into the posterior chamber underneath the iris. After optimizing the process, the device was successfully implanted in three eyes of three living phakic primates. An inflammatory response was visible during the postoperative follow-up period, but postmortem histopathological analysis revealed no clinically relevant adverse changes.

CONCLUSIONS

The primate model demonstrated the feasibility of the surgical implantation of a circular electrode into the posterior chamber. Regarding future developments of biomimetic accommodative IOLs, a concomitant improvement of surgical techniques and the device is crucial to ensure treatment success.

Для цитирования: Kortuem Friederike C, Seitz Immanuel, Kaltenstadler Sebastian, Schumayer Sven, Sigdel Bishesh, Peters Tobias, Rothermel Albrecht, Bucher Volker, Zrenner Eberhart, Bartz-Schmidt Karl-Ulrich, Straßer Torsten. Surgical Approach for the Implantation of a Ring Electrode into the Sulcus in a Phakic Non-human Primate.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2026. DOI: 10.3928/1081597x-20251217-02.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление