BACKGROUND

Reverse shoulder arthroplasty (RSA) is commonly used to treat patients with rotator cuff tear arthropathy. Loosening of the glenoid component remains one of the principal modes of failure and represents a significant complication that requires revision surgery. This study assessed the effects of various factors on glenoid baseplate micromotion for primary fixation of RSA.

MATERIALS AND

METHODS

A half-fractional factorial design of experiment was used to assess 4 factors: central element type (central peg or screw), central cortical engagement according to length (13.5 or 23.5 mm), anterior-posterior peripheral screw type (nonlocking or locking), and cancellous bone surrogate density (160 or 400 kg/m3, 10 or 25 PCF). Glenoid baseplates were implanted into high- or low-density Sawbones rigid polyurethane foam blocks and cyclically loaded at 60° for 1000 cycles (500-N compressive force range) using a custom-designed loading apparatus. Micromotion at the 4 peripheral screw positions was recorded using linear variable differential transformers.

RESULTS

Central peg fixation generated 358% greater micromotion at all peripheral screw positions compared with central screw fixation (P < .001). Baseplates with short central elements that lacked cortical bone engagement generated 328% greater micromotion than those with long central elements (P = .001). No significant effects were observed when varying anterior-posterior peripheral screw type or bone surrogate density. There were significant interactions between central element type and length (P < .001).

DISCUSSION: A central screw and a long central element that engaged cortical bone reduced RSA baseplate micromotion. These findings serve to inform surgical decision making regarding baseplate fixation elements to minimize the risk of glenoid loosening and, thus, the need for revision surgery.

Для цитирования: Torkan Lawrence F, Bryant John T, Bicknell Ryan T, Ploeg Heidi-Lynn. Central fixation element type and length affect glenoid baseplate micromotion in reverse shoulder arthroplasty.. Journal of shoulder and elbow surgery. 2022. DOI: 10.1016/j.jse.2022.01.120.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.