DOI 10.1007/s00417-016-3537-1

Медицинская библиотека · Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2016

Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery.

López-Miguel Alberto, Sanchidrián María, Fernández Itziar, Holgueras Alfredo, Maldonado Miguel J

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To compare specular microscopy (Topcon SP-3000P, Topcon Corporation, Tokyo, Japan) and ultrasound (US) technology when evaluating central corneal thickness (CCT) prior to and after phacoemulsification.

METHODS

Corneal edema was assessed in phacoemulsification patients due to senile cataract by measuring CCT preoperatively and 1 day, 1 month, and 3 months postoperatively. Bland-Altman analysis was performed to assess interchangeability between pachymetry techniques for each visit. Repeated measures analysis of variance was performed to evaluate variation in CCT agreement depending on the degree of corneal edema.

RESULTS

One hundred and eighteen patients aged 73.9 ± 10.1 years were recruited. Topcon SP-3000P provided significant (p < 0.0001) lower CCT values than US with and without induced corneal edema. Mean differences between CCT techniques for baseline, 1 day, 1 month and 3 months after cataract surgery were -28.9 ± 22.6, -30.5 ± 41.4, -32.3 ± 16.2 and -33.0 ± 16.9 μm, respectively. The systematic bias observed was not significantly different among the 4 visits (p = 0.59). The estimated limits of agreement (based on 1.96 standard deviation) were substantial, being 90.5, 165.8, 64.9 and 67.5 um at baseline, 1 day and 1 and 3 months.

CONCLUSIONS

Topcon SP-3000P provides similar systematic bias in comparison with US technique for CCT measurements regardless of the degree of corneal edema after phacoemulsification, which should be reduced after applying a constant calibration adjustment of ∼30 μm. However, interchangeability between techniques may be still limited by the notable random measurement error.

Для цитирования: López-Miguel Alberto, Sanchidrián María, Fernández Itziar, Holgueras Alfredo, Maldonado Miguel J. Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery.. Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie. 2016. DOI: 10.1007/s00417-016-3537-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.