Journal of glaucoma · 2021

Variations of Intraocular Pressure Measured by Goldmann Applanation Tonometer, Tono-Pen, iCare Rebound Tonometer, and Pascal Dynamic Contour Tonometer in Patients With Corneal Edema After Phacoemulsification.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kiddee Weerawat, Tanjana Arunee

Аннотация

PRCIS: Edematous corneas had lower Goldmann applanation tonometer (GAT) intraocular pressure (IOP) compared with other tonometers. A significant, mild negative correlation between central corneal thickness (CCT) and GAT IOP was found in a group of significant edematous corneas with CCT increases of >100 µm.

PURPOSE: To compare the IOP that was randomly obtained with the GAT, Tono-Pen, iCare, and Pascal dynamic contour tonometer in patients with corneal edema after phacoemulsification.

MATERIALS AND METHODS: Corneal edema was quantified by CCT measurement. The agreement between IOP measurements assessed by different instruments was evaluated using Bland-Altman plots. The relationship between CCT and IOP was investigated using the coefficient of correlation. The CCT and IOP were compared between the periods of postsurgical edema and after its resolution.

RESULTS: The mean age of 60 patients was 75.9±7.8 years. CCT increased significantly after surgery, by an average of 96.1±39.9 µm (P<0.001). Relative to the IOPs measured with the GAT in edematous eyes: the mean difference in Tono-Pen IOPs was 4.7 mm Hg; 95% limits of agreement (LOA), -0.3 to 9.6 mm Hg. For iCare, was 2.3 mm Hg; 95% LOA, -2.2 to 6.8 mm Hg. For Pascal dynamic contour tonometer, was 3.0 mm Hg; 95% LOA, -0.4 to 6.5 mm Hg. Edematous corneas had significantly lower GAT IOP than Tono-Pen IOP; 12.3±3.2 versus 16.9±3.1 (P=0.02). Pearson correlation coefficients (r) showed a high correlation between the 4 tonometers in eyes without edema. There was a significant, mild negative correlation between CCT and GAT IOP (r=-0.25, P=0.02) in a group of significant edematous corneas with CCT increases of >100 µm: GAT IOP decreased by 3.1 mm Hg for every 10-µm increase in CCT.

CONCLUSIONS: IOP readings with GAT tended to be lower than those obtained with the other tonometers, especially the Tono-Pen. GAT IOP readings in cases of severe corneal edema should be interpreted with caution.

Предметные рубрики MeSH