DOI 10.4293/jsls.2020.00031

Медицинская библиотека · JSLS : Journal of the Society of Laparoendoscopic Surgeons · 2020

Impact of Refractive Errors on Da Vinci SI Robotic System.

Tuna Mustafa Bilal, Kilavuzoglu Ayse Ebru, Mourmouris Panogiotis, Argun Omer Burak, Doganca Tunkut, Obek Can, Ozisik Ozan, Kural Ali Riza

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To investigate the impact of refractive errors on binocular visual acuity while using the Da Vinci SI robotic system console.

METHODS

Eighty volunteers were examined on the Da Vinci SI robotic system console by using a near vision chart. Refractive errors, anisometropia status, and Fly Stereo Acuity Test scores were recorded. Spherical equivalent (SE) were calculated for all volunteers' right and left eyes. Visual acuity was assessed by the logarithm of the minimal angle of resolution (LogMAR) method. Binocular uncorrected and best corrected (with proper contact lens or glasses) LogMAR values of the subjects were recorded. The difference between these values (DiffLogMAR) are affected by different refractive errors.

RESULTS

In the myopia and/or astigmatism group, uncorrected SE was found to have significant impact on the DiffLogMAR (p < 0.001) and myopia greater than 1.75 diopter had significantly higher DiffLogMAR values (p < 0.05). Subjects with presbyopia had significantly higher DiffLogMAR values (p < 0.01), and we observed positive correlation between presbyopia and DiffLogMAR values (p = 0.33, p < 0.01). The cut off value of presbyopia that correlated the most with DiffLogMAR differences was found to be 1.25 diopter (p < 0.001). In 13 hypermetropic volunteers, we found significant correlation between hypermetropia value and DiffLogMAR (p > 0.7, p < 0.01). The statistical analysis between Fly test and SE revealed a significant impact of presbyopia and hypermetropia to the stereotactic view of the subject (p = -0.734, p < 0.05).

CONCLUSION

Surgeons suffering from myopia greater than 1.75 diopter, presbyopia greater than 1.25 diopter (D), and hypermetropia regardless of grade must always perform robotic surgeries with the proper correction.

Для цитирования: Tuna Mustafa Bilal, Kilavuzoglu Ayse Ebru, Mourmouris Panogiotis, Argun Omer Burak, Doganca Tunkut, Obek Can, Ozisik Ozan, Kural Ali Riza. Impact of Refractive Errors on Da Vinci SI Robotic System.. JSLS : Journal of the Society of Laparoendoscopic Surgeons. 2020. DOI: 10.4293/jsls.2020.00031.

Предметные рубрики MeSH