Varus-valgus alignment of humeral short stem in reverse total shoulder arthroplasty: does it really matter?

Cho Sung-Hyun, Park Chan-Joo, Kim Sang-Jae, Lee Kyoung-Geun, Baek Gyu Rim, Chung Min-Shik, Hui Aaron T, McGarry Michelle H, Lee Thay Q, Jung Sungwook, Kim Jaewon, Kim Yang-Soo

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The utilization of short humeral stems in reverse total shoulder arthroplasty has gained attention in recent times. However, concerns regarding the risk of misalignment during implant insertion are associated with their use.

METHODS

Eight fresh-frozen cadaveric shoulders were prepared for dissection and biomechanical testing. A bespoke humeral implant was fabricated to facilitate assessment of neutral, varus, and valgus alignments using a single stem, and 10° was established as the maximum permissible angle for misalignments. Shift in humerus position and changes in deltoid length attributable to misalignments relative to the neutral position were evaluated using a Microscribe 3DLx system. The impingement-free range of motion, encompassing abduction, adduction, internal rotation, and external rotation (ER), was gauged using a digital goniometer. The capacity for abduction was evaluated at maximal abduction angles under successive loading on the middle deltoid. A specialized traction system coupled with a force transducer was employed to measure anterior dislocation forces.

RESULTS

Relative to the neutral alignment, valgus alignment resulted in a more distal (10.5 ± 2.4 mm) and medial (8.3 ± 2.2 mm) translation of the humeral component, whereas the varus alignment resulted in the humerus shifting more superiorly (11.2 ± 1.3 mm) and laterally (9.9 ± 0.9 mm) at 0° abduction. The valgus alignment exhibited the highest abduction angle than neutral alignment (86.2°, P < .001). Conversely, the varus alignment demonstrated significantly higher adduction (18.4 ± 7.4°, P < .001), internal rotation (68.9 ± 15.0°, P = .014), and ER (45.2 ± 10.5°, P = .002) at 0° abduction compared to the neutral alignments. Anterior dislocation forces were considerably lower (23.8 N) in the varus group compared to the neutral group at 0°ER (P = .047). Additionally, abduction capability was markedly higher in varus alignment at low deltoid loads than the neutral alignment (5N, P = .009; 7.5 N, P = .007).

CONCLUSIONS

The varus position enhances rotational range of motion (ROM) but increases instability, while the valgus position does not significantly impact ROM or instability compared to the neutral position.

Для цитирования: Cho Sung-Hyun, Park Chan-Joo, Kim Sang-Jae, Lee Kyoung-Geun, Baek Gyu Rim, Chung Min-Shik, Hui Aaron T, McGarry Michelle H, Lee Thay Q, Jung Sungwook, Kim Jaewon, Kim Yang-Soo. Varus-valgus alignment of humeral short stem in reverse total shoulder arthroplasty: does it really matter?. Journal of shoulder and elbow surgery. 2024. DOI: 10.1016/j.jse.2023.11.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.