Satou Tsukasa, Kato Sayaka, Igarashi Akihito, Ito Misae, Tsunehiro Shuntaro, Koshimizu Manabu, Niida Takahiro, Shimizu Kimiya
Аннотация
PURPOSE: Pupillometry should be performed under conditions as close to natural viewing as possible. The present study aimed to determine whether pupil size in binocular open-view settings can be predicted based on pupil size measured using the CASIA2 device.
METHODS: The present study included 61 participants (25 men and 36 women; mean age, 49 ± 15 years; age range, 22-69 years) with no history of ophthalmic disease other than refractive errors and cataract. We measured pupil size using the new CASIA2 device and a binocular open-view digital pupillometer (FP-10000II, TMI Co., Ltd., Saitama). Intra-class and inter-class reliabilities were evaluated by measuring pupil times three times with each device (two independent examiners) in 21 of the 61 participants. Reproducibility was analyzed using intra-class and inter-class correlation coefficients (ICCs). Regression formulae for calculating FP10000II pupil size based on CASIA2 pupil size were developed via simple linear regression analyses.
RESULTS: Both devices exhibited high ICC values (> 0.80). The regression formulae for calculating the FP10000II pupil size for the distant and near views based on CASIA2 pupil size were y = 0.5702x + 0.4611 (determination coefficient, 0.67) and y = 0.502x + 0.445 (determination coefficient, 0.64), respectively.
CONCLUSIONS: Pupil size under binocular open-view settings can be predicted based on simultaneous measurement of pupil size during evaluation of the anterior segment using the CASIA2 device. The calculated pupil size may represent a useful index for determining the most appropriate treatment strategy in candidates for cataract and refractive surgery.
METHODS: The present study included 61 participants (25 men and 36 women; mean age, 49 ± 15 years; age range, 22-69 years) with no history of ophthalmic disease other than refractive errors and cataract. We measured pupil size using the new CASIA2 device and a binocular open-view digital pupillometer (FP-10000II, TMI Co., Ltd., Saitama). Intra-class and inter-class reliabilities were evaluated by measuring pupil times three times with each device (two independent examiners) in 21 of the 61 participants. Reproducibility was analyzed using intra-class and inter-class correlation coefficients (ICCs). Regression formulae for calculating FP10000II pupil size based on CASIA2 pupil size were developed via simple linear regression analyses.
RESULTS: Both devices exhibited high ICC values (> 0.80). The regression formulae for calculating the FP10000II pupil size for the distant and near views based on CASIA2 pupil size were y = 0.5702x + 0.4611 (determination coefficient, 0.67) and y = 0.502x + 0.445 (determination coefficient, 0.64), respectively.
CONCLUSIONS: Pupil size under binocular open-view settings can be predicted based on simultaneous measurement of pupil size during evaluation of the anterior segment using the CASIA2 device. The calculated pupil size may represent a useful index for determining the most appropriate treatment strategy in candidates for cataract and refractive surgery.