Contact dermatitis caused by glucose sensors-15 adult patients tested with a medical device patch test series.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ulriksdotter Josefin, Svedman Cecilia, Bruze Magnus, Glimsjö Jenny, Källberg Kajsa, Sukakul Thanisorn, Mowitz Martin
Аннотация
BACKGROUND: Several cases of allergic contact dermatitis (ACD) to the glucose sensor FreeStyle Libre have been reported. Isobornyl acrylate (IBOA) and N,N-dimethylacrylamide (DMAA) are known culprit allergens.
OBJECTIVES: To evaluate patients with suspected ACD to FreeStyle Libre in a standardized manner, present causative allergens, and assess patient-reported implications.
METHODS: A total of 15 patients with suspected ACD to FreeStyle Libre were patch tested with the Swedish baseline series and a new medical device series. IBOA and DMAA were tested at 0.1% and 0.3% in petrolatum (pet.). Readings were performed on day (D) 3 and D7. Background data, details on skin reactions, and associated implications were assessed using a questionnaire.
RESULTS: Thirteen patients were sensitized to IBOA and four to DMAA. Two positive reactions to IBOA and one to DMAA were seen only at 0.3% concentration on D7. Median duration of sensor use before dermatitis onset was 6 months. Half the number of the patients took precautions in everyday life due to sensor-related skin reactions. Six patients discontinued sensor usage.
CONCLUSIONS: Patients with suspected ACD to glucose sensors should be evaluated with a relevant patch test series containing IBOA and DMAA. Adding the 0.3% pet. concentration is recommended. The reading on D7 is necessary.
OBJECTIVES: To evaluate patients with suspected ACD to FreeStyle Libre in a standardized manner, present causative allergens, and assess patient-reported implications.
METHODS: A total of 15 patients with suspected ACD to FreeStyle Libre were patch tested with the Swedish baseline series and a new medical device series. IBOA and DMAA were tested at 0.1% and 0.3% in petrolatum (pet.). Readings were performed on day (D) 3 and D7. Background data, details on skin reactions, and associated implications were assessed using a questionnaire.
RESULTS: Thirteen patients were sensitized to IBOA and four to DMAA. Two positive reactions to IBOA and one to DMAA were seen only at 0.3% concentration on D7. Median duration of sensor use before dermatitis onset was 6 months. Half the number of the patients took precautions in everyday life due to sensor-related skin reactions. Six patients discontinued sensor usage.
CONCLUSIONS: Patients with suspected ACD to glucose sensors should be evaluated with a relevant patch test series containing IBOA and DMAA. Adding the 0.3% pet. concentration is recommended. The reading on D7 is necessary.