MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/mrm.70089

Медицинская библиотека · Magnetic resonance in medicine · 2025

Focus correction in MR thermography for increased targeting precision during focused ultrasound procedures.

Mei Chang-Sheng, Zong Shenyan, Madore Bruno, Cosgrove G Rees, McDannold Nathan J

СтатьяПеревод в очереди

Аннотация

PURPOSE: Accurate targeting during MR-guided focused ultrasound (FUS) procedures is essential for effective treatment to be achieved. However, spatial discrepancies frequently arise between the planned target and the observed thermal hotspot on proton resonance frequency (PRF)-based MR thermometry because of temperature-induced artifacts. This study aims to correct such displacements caused by chemical shift and k-space center offset.

METHODS

Spatial misregistration was addressed using a two-step correction approach. The first step corrected pixel-wise displacements attributed to temperature-dependent resonance frequency shifts (chemical shift), based on local frequency offset maps. The second step compensated for TE errors induced by asymmetric phase gradients near the thermal focus, restoring accuracy in hotspot localization. Validation was performed in controlled phantom experiments, and the approach was retrospectively tested in vivo, in 121 sonications across seven essential tremor (ET) patients.

RESULTS

Phantom experiments demonstrated that spatial shifts up to approximately 1.5 mm could be effectively corrected. Clinical analysis showed a strong correlation (R2 = 0.852) between temperature rise and spatial displacement, with a mean shift of 0.5 Mm per 10°C. Combined correction significantly reduced temperature estimation bias, with the mean error decreasing from -0.11°C to -0.05°C, as evaluated by Bland-Altman analysis.

CONCLUSION

Temperature-related chemical shift and k-space offset substantially impact the spatial fidelity of PRF-based MR thermometry. The proposed correction framework improves thermal hotspot localization, enabling more accurate lesion targeting during FUS procedures.

Для цитирования: Mei Chang-Sheng, Zong Shenyan, Madore Bruno, Cosgrove G Rees, McDannold Nathan J. Focus correction in MR thermography for increased targeting precision during focused ultrasound procedures.. Magnetic resonance in medicine. 2025. DOI: 10.1002/mrm.70089.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.