MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2147/imcrj.s529596

Медицинская библиотека · International medical case reports journal · 2025

Right Internal Jugular Vein Access for Micra Leadless Pacemaker Implantation in a High-Risk Patient with Severe Iliac Vein Stenosis: A Case Report.

Xu Haixiang, Qian Jianfeng, Pan Wen, Fan Jianhua

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

Leadless pacemakers offer a promising alternative to traditional transvenous pacemakers, especially in patients with complex vascular anatomy or a high risk of infection. Conventional access routes may be compromised in patients with significant venous stenosis or prior vascular interventions, requiring alternative strategies for device implantation.

CASE PRESENTATION: We report the case of a 74-year-old female with sick sinus syndrome, chronic kidney disease, type 2 diabetes, and severe bilateral iliac vein stenosis, who was deemed high-risk for conventional pacemaker implantation. After failed attempts through both femoral veins, the right internal jugular vein (RIJV) was successfully used as an alternative access route for Micra leadless pacemaker implantation. The procedure was performed under ultrasound and fluoroscopic guidance without complications. Post-procedure, the pacing threshold was 0.75 V, sensing was 8.5 mV, and impedance was 760 Ω, all within normal limits.

CONCLUSION

This case highlights the feasibility and safety of using the right internal jugular vein as an alternative access route for leadless pacemaker implantation in high-risk patients when femoral access is not possible. This approach provides a valuable option for patients with complex venous anatomy, contributing to the broader application of leadless pacing technology.

Для цитирования: Xu Haixiang, Qian Jianfeng, Pan Wen, Fan Jianhua. Right Internal Jugular Vein Access for Micra Leadless Pacemaker Implantation in a High-Risk Patient with Severe Iliac Vein Stenosis: A Case Report.. International medical case reports journal. 2025. DOI: 10.2147/imcrj.s529596.