DOI 10.1111/j.1540-8167.2006.00686.x
Transvenous cardioverter-defibrillator implantation in a patient with tricuspid mechanical prosthesis.
Аннотация
BACKGROUND
A 64-year-old woman was referred to our center because of poorly tolerated ventricular tachycardia (VT) at 210 bpm due to an old myocardial infarction. The patient had been operated on at age of 20 for mitral valve commissurolysis, at age of 49 for ductal carcinoma, at age of 56 for mitral valve replacement, and at age of 61 for tricuspid valve replacement. Left ventricular EF was 31%. The patient was in permanent atrial fibrillation (AF) since the age of 53. She had undergone three cardiac surgery procedures, ending with two prosthetic mechanical valves. The cardiac surgery team advised against an epicardial ICD implantation.
RESULTS
We achieved a fully transvenous implant, with a screw-in defibrillation coil in the low right atrium and a bipolar pacing/sensing lead in a posterolateral branch of the coronary sinus. Pacing/sensing parameters were reliable, and effective defibrillation occurred at 20 J by a stepdown protocol. During 16-month follow-up, three VT episodes at 210 bpm were terminated by antitachycardia pacing (ATP) therapy. Left ventricular pacing/sensing was stable at long term.
CONCLUSION
Thanks to technologic improvements, transvenous ICD implantation is feasible and safe in patients with a tricuspid mechanical prosthesis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.