Transvenous cardioverter-defibrillator implantation in a patient with tricuspid mechanical prosthesis.

Biffi Mauro, Bertini Matteo, Ziacchi Matteo, Boriani Giuseppe

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

A 64-year-old woman was referred to our center because of poorly tolerated ventricular tachycardia (VT) at 210 bpm due to an old myocardial infarction. The patient had been operated on at age of 20 for mitral valve commissurolysis, at age of 49 for ductal carcinoma, at age of 56 for mitral valve replacement, and at age of 61 for tricuspid valve replacement. Left ventricular EF was 31%. The patient was in permanent atrial fibrillation (AF) since the age of 53. She had undergone three cardiac surgery procedures, ending with two prosthetic mechanical valves. The cardiac surgery team advised against an epicardial ICD implantation.

RESULTS

We achieved a fully transvenous implant, with a screw-in defibrillation coil in the low right atrium and a bipolar pacing/sensing lead in a posterolateral branch of the coronary sinus. Pacing/sensing parameters were reliable, and effective defibrillation occurred at 20 J by a stepdown protocol. During 16-month follow-up, three VT episodes at 210 bpm were terminated by antitachycardia pacing (ATP) therapy. Left ventricular pacing/sensing was stable at long term.

CONCLUSION

Thanks to technologic improvements, transvenous ICD implantation is feasible and safe in patients with a tricuspid mechanical prosthesis.

Для цитирования: Biffi Mauro, Bertini Matteo, Ziacchi Matteo, Boriani Giuseppe. Transvenous cardioverter-defibrillator implantation in a patient with tricuspid mechanical prosthesis.. Journal of cardiovascular electrophysiology. 2007. DOI: 10.1111/j.1540-8167.2006.00686.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.