Analytical and clinical performance of different platforms simultaneously detecting 15 antinuclear antibodies.

Qin Yan, Fan Chunxue, Wang Yanlin, Feng Min, Liang Zhaojun, Zhao Xiangcong, Gao Chong, Luo Jing

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Antinuclear antibodies (ANAs) are invaluable biomarkers for the diagnosis of autoimmune diseases (AIDs). This study aims to compare the performances of line immunoassay (LIA), multiplex bead-based flow fluorescent immunoassay (MBFFI), and magnetic bar code immunofluorescence assay (MBC-IF) to detect ANA-Profile-15S.

METHODS

In total, 184 samples from AID patients and 50 healthy controls (HCs) were collected. Fifteen ANAs (anti-dsDNA, nucleosome, histone, Sm, PCNA, ribosomal-P, SS-A/Ro52, SS-A/Ro60, SS-B/La, centromere B [CENP-B], Scl-70, U1-snRNP, AMA-M2, Jo-1, and Pm/Scl) were subjected to parallel detection by the LIA, MBFFI, and MBC-IF. The consistency between assays was analyzed. The discrepant results were further examined by chemiluminescent immunoassay (CLIA).

RESULTS

Anti-SS-A/Ro52 and SS-A/Ro60 autoantibodies were the most common autoantibodies in ANA positive-profiles, and were detected with equal efficiency by the LIA, MBFFI, and MBC-IF (p = 0.101 and p = 0.732, respectively). The three assays showed excellent agreement (consistency range: 66.5%-97.5%), and total consistency was 85.8%. The MBFFI and MBC-IF assays were in good agreement in terms of ANA-Profile-15S determination; the kappa coefficient ranged from 0.59 to 0.95, except for the PCNA and PM-Scl. Of the 262 re-assessed divergent results, 124 (47.33%) were positive on CLIA; the various autoantibodies exhibited variable patterns. More importantly, the ANA-Profile-15S results of the MBFFI and MBC-IF accurately identified patients with AID; the area under the curves ranged from 0.642 to 0.919.

CONCLUSIONS

The novel MBFFI and MBC-IF assay performed well in detecting ANA-Profile-15S. The application of MBFFI and MBC-IF play important roles in laboratory diagnosis of AIDs.

Для цитирования: Qin Yan, Fan Chunxue, Wang Yanlin, Feng Min, Liang Zhaojun, Zhao Xiangcong, Gao Chong, Luo Jing. Analytical and clinical performance of different platforms simultaneously detecting 15 antinuclear antibodies.. Journal of clinical laboratory analysis. 2022. DOI: 10.1002/jcla.24554.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.