MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.20841

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2006

Comparison of transendocardial and retrograde coronary venous intramyocardial catheter delivery systems in healthy and infarcted pigs.

Baklanov D V, Moodie K M, McCarthy F E, Mandrusov E, Chiu J, Aswonge G, Cheng J, Chow M, Simons M, de Muinck E D

СтатьяПеревод в очереди

Аннотация

UNLABELLED: We compared two routes for myocardial delivery of therapeutics, transendocardial (TE) delivery with an intramyocardial injection catheter, and retrograde coronary venous (RCV) delivery with a balloon occlusion catheter in the interventricular vein.

METHODS

TE and RCV injection of 15 microM, neutron-activatable microspheres was compared in healthy pigs (Group I, n = 3), pigs with a 1-week-old myocardial infarction (MI; group II, n = 5), and pigs with a 2-weeks-old MI (group III, n = 4). The MI was induced by a 1-hr balloon occlusion in the LAD. Both methods were compared in the same animal using different microspheres. The RCV catheter allowed for continuous measurement of distal pressure and 2.5 x 10(6) microspheres were injected in 10 ml at 300 mmHg above balloon occlusion pressure. The TE injections were targeted to the infarct zone and 2.5 x 10(6) microspheres were distributed over 10 injections of 200 microl.

RESULTS

The retention of microspheres decreased with increase in MI age, but was comparable between devices within the groups. RCV delivery resulted in (14.3 +/- 0.9)% microsphere retention in Group I, (10.3 +/- 0.2)% in Group II, and (6.4 +/- 0.1)% in group III (P < 0.05 versus group I). Microsphere retention after TE was (15.1 +/- 0.7)% in group I, (18.9 +/- 0.6)% in group II, (4.1 +/- 0.1)% in Group III (P < 0.05 versus groups I and II). The RCV catheter delivered primarily to midventricular, antero-septal segments, whereas TE targeted apical areas predominantly.

CONCLUSIONS

Delivery efficacy was comparable between devices in each group however RCV targeted midventricular areas whereas TE targeted apical areas.

Для цитирования: Baklanov D V, Moodie K M, McCarthy F E, Mandrusov E, Chiu J, Aswonge G, Cheng J, Chow M, Simons M, de Muinck E D. Comparison of transendocardial and retrograde coronary venous intramyocardial catheter delivery systems in healthy and infarcted pigs.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2006. DOI: 10.1002/ccd.20841.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.