Is invasive fractional flow measurement accurate in intracranial stenosis? A computational simulation study.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Liu Yu, Yan Zhengzheng, Li Ziqi, Liu Yuying, Ma Sze Ho, Ip Bonaventure Yiu Ming, Leung Thomas Wai Hong, Liu Jia, Leng Xinyi

Аннотация

BACKGROUND: In intracranial atherosclerotic stenosis (ICAS), low fractional flow (FF) may indicate hemodynamic significance.

OBJECTIVE: To investigate, using simulation models, whether invasive measurement could accurately reflect 'true' FF, when the catheter/pressure wire might disturb focal flow during measurement.

METHODS: We recruited 5 patients with high-grade, symptomatic M1 middle cerebral artery stenosis in three-dimensional rotational angiography (3DRA). In each case, the stenotic severity was manually manipulated to 50%, 60%, 70%, and 80%. At each stenotic severity, we simulated four situations: no catheter/pressure wire insertion ('in vivo' status); only catheter inserted proximally to the ICAS; catheter inserted and pressure sensor placed at 1 cm and 2 cm distally to the ICAS lesion. The blood flow was simulated with computational fluid dynamics modeling, and FF measured as post-stenotic and pre-stenotic pressure ratio. We calculated relative differences of FFs simulated at other situations compared with in vivo status.

RESULTS: Compared with in vivo status, catheter insertion had no significant influence on simulated FFs. With pressure wire passing through the ICAS, simulated FFs slightly decreased (mostly <10%) at 50% and 60% stenoses, which significantly decreased (up to 50% and 88%) at 70% and 80% stenoses. The effects of pressure wire on FFs were similar when the pressure sensor was placed at 1 cm and 2 cm distally to the ICAS.

CONCLUSIONS: Invasive measurement of FF may overestimate the hemodynamic significance of ICAS with severe stenosis, as the pressure wire may further reduce the flow across the small residual lumen. The findings warrant verification in larger-scale studies, with information on collateral circulation and validation with other imaging modalities.