MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcjo.2025.08.012

Медицинская библиотека · Canadian journal of ophthalmology. Journal canadien d'ophtalmologie · 2025

Optimization of ciliary body cyclophotocoagulation settings: a randomized and controlled pilot study.

Carrière Mathieu, Antoniades Paul-Alexandre Pavlos, Marchais Béatrice Des, Toren A

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To compare the efficacy of different settings in continuous-wave transscleral cyclophotocoagulation (CW-TSCPC).

DESIGN

Prospective, randomized, triple-blind clinical trial in a single tertiary center.

PARTICIPANTS

Adult glaucoma patients who need continuous wave-transscleral cyclophotocoagulation to control intraocular pressure (IOP) or prevent or treat ocular pain/discomfort secondary to increased IOP.

METHODS

Group A uses 1 250 mW for 4 seconds. Group B uses 2 000 mW for 2 seconds.

MAIN OUTCOME MEASURES: IOP, best-corrected visual acuity, glaucoma medication, complications, and the need to retreat.

RESULTS

Eighteen and 19 eyes in group A and group B, respectively, were included. Groups were comparable at baseline (p > 0.1) for mean age (76.9, 73.5 years), mean IOP (29.1, 29.8 mm Hg), mean logMAR visual acuity (1.6, 1.7), and mean number of glaucoma agents (3.4, 3.4). IOP was comparable between groups at 1 (15.6, 13.7 mm Hg), 4-6 (14.6, 14.8 mm Hg) and 12 months (10.3, 17.3 mm Hg; p = 0.07). There was a trend toward more medication reduction earlier in group B at 1 and 4-6 months (p = 0.11). Complication rates were similar between groups (group A: n = 6; group B: n = 9), with more serious complications in group B. Reintervention was required in 5 patients in group A, and 3 patients in group B after a mean of 4.7 and 6.7 months, respectively.

CONCLUSIONS

Both techniques are efficient and comparable in reducing IOP. Our study favours the 1 250 mW for 4 seconds group, with its possibly better safety profile.

Для цитирования: Carrière Mathieu, Antoniades Paul-Alexandre Pavlos, Marchais Béatrice Des, Toren A. Optimization of ciliary body cyclophotocoagulation settings: a randomized and controlled pilot study.. Canadian journal of ophthalmology. Journal canadien d'ophtalmologie. 2025. DOI: 10.1016/j.jcjo.2025.08.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.