DOI 10.1097/j.jcrs.0000000000000978
Clinical outcomes in a U.S. registration study of a new EDOF intraocular lens with a nondiffractive design.
Аннотация
PURPOSE: To evaluate the effectiveness and safety of the DFT015 intraocular lens (IOL) (AcrySof IQ Vivity Extended Vision) compared with an aspheric monofocal control IOL (AcrySof IQ model SN60WF).
SETTING
11 investigation sites in the U.S.
DESIGN
Prospective randomized controlled clinical study.
METHODS
Patients aged 22 years or older with bilateral cataracts were randomized to receive bilateral implantation of DFT015 or SN60WF. The 4 coprimary effectiveness outcomes (6 months postoperatively) were monocular photopic distance-corrected intermediate visual acuity (DCIVA), monocular photopic corrected distance visual acuity (CDVA), monocular depth of focus (DoF), and the percentage of patients achieving a DCIVA of 0.2 logMAR or better. The mean monocular photopic distance-corrected near visual acuity (DCNVA) was a secondary effectiveness outcome. Safety and patient-reported visual disturbances were evaluated through questionnaires.
RESULTS
218 patients (435 eyes) completed the study. Compared with SN60WF, DFT015 demonstrated superior mean monocular photopic DCIVA ( P < .001), noninferior mean monocular photopic CDVA, and superior mean monocular photopic DCNVA ( P < .001) and provided an extended monocular DoF (increase of 0.54 diopters at 0.2 logMAR). With DFT015, 78 first eyes (72.9%) achieved a DCIVA of 0.2 logMAR or better at 6 months. Incidences of ocular serious adverse events and patient-reported most bothersome visual disturbances were low and consistent between groups.
CONCLUSIONS
DFT015 is safe and effective for the visual correction of aphakia, exceeding American National Standards Institute criteria for an extended depth-of-focus IOL by providing superior DCIVA and DCNVA, with comparable CDVA and visual disturbances to the SN60WF monofocal IOL.