DOI 10.1002/jcla.70032

Медицинская библиотека · Journal of clinical laboratory analysis · 2025

Comparison of Enzyme-Linked Immunosorbent Assay and Lateral Flow Assay to Measure Thyroid-Stimulating Hormone and Free T4 in Human Serum.

Advíncula-Espino Rolando, Li Jair, Rosales-Rimache Jaime

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Thyroid disorders affect millions of people around the world, mainly women. Measurement of hormones is critical in their identification, treatment, and monitoring; Therefore, the methods must be as reliable and valid as possible.

METHODS

We designed a cross-sectional study to compare thyroid-stimulating hormone (TSH) and free tetraiodothyronine (fT4) concentrations using ELISA (Enzyme-Linked Immunosorbent Assay) and LFA (Lateral Flow Assay) in human serum samples. We obtained 96 serum samples that were evaluated with a commercial kit for ELISA (Accubind Monobite, Lake Forest, California, US) and LFA (Micropofit Fluorecare, Guangdong, Shenzhen, China). We used the Wilcoxon nonparametric test, Bland-Altman, and Passing-Bablok regression to compare TSH and fT4 concentrations obtained by ELISA and LFA.

RESULTS

The median TSH concentrations obtained by ELISA and LFA were 1.92 and 2.11 μIU/mL, and for fT4 were 1.14 and 1.10 ng/dL. On the other hand, the Spearman's rho between ELISA and LFA for TSH was 0.845. The TSH concentrations between ELISA and LFA had significant differences (p < 0.05). About fT4, the Spearman's rho between methods was 0.348. The bias for TSH was -0.315, and for fT4 was -0.013. The Deming regression for TSH (p = 0.309) and fT4 (=0.938) shows that the levels obtained by both methods do not present significant differences, and the Passing-Bablok regression identifies significant bias between both methods, especially for fT4, in the range of concentrations studied.

CONCLUSION

TSH measurement by LFA may be a viable alternative to evaluate thyroid diseases, but fT4 by LFA is not precise and presents a high bias.

Для цитирования: Advíncula-Espino Rolando, Li Jair, Rosales-Rimache Jaime. Comparison of Enzyme-Linked Immunosorbent Assay and Lateral Flow Assay to Measure Thyroid-Stimulating Hormone and Free T4 in Human Serum.. Journal of clinical laboratory analysis. 2025. DOI: 10.1002/jcla.70032.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.