Smartphone-Based Pupillometry for Noninvasive Detection of Elevated Intracranial Pressure in Nepali Acute Brain Injury Patients: A Pilot Study.

Pant Aaradhya, Maxin Anthony, Farrokhi Farrokh, Shrestha Gentle S, Sarwal Aarti, Bohara Sandeep, Veeravagu Anand, Sharma Mohan R

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Intracranial hypertension is a life-threatening complication of acute brain injuries such as traumatic brain injury (TBI), subarachnoid hemorrhage (SAH), or intracerebral hemorrhage (ICH). In low-income and middle-income countries (LMICs), limited resources can delay timely neurocritical interventions. Smartphone-based quantitative pupillometry offers a scalable solution for early detection of elevated intracranial pressure (ICP). Here, we assessed its ability to (1) detect raised optic nerve sheath diameter (ONSD), a noninvasive surrogate for elevated ICP, and (2) classify severe TBI.

METHODS

Thirty-eight Nepali ICU patients with TBI (n=16), SAH (n=10), or ICH (n=12) underwent daily sonographic ONSD and pupillary light reflex (PLR) assessments through the PupilScreen app (Apertur Inc., Seattle, WA) over 7 days. Machine learning classifiers were trained on PLR features to detect elevated ONSD (>6.0 mm). To identify severe TBI (Glasgow Coma Scale [GCS] ≤8 on admission), classifiers were trained on PLR features, ONSD, or both.

RESULTS

For ONSD >6.0 mm, a random forest model achieved an AUC of 0.66, with a sensitivity of 0.31 and specificity of 0.80. For identifying severe TBI, the optimal classifier was a random forest model incorporating ONSD and a subset of PLR metrics, with a sensitivity of 0.93, specificity of 1.00, and AUC of 0.96.

CONCLUSION

In this pilot study, smartphone-based pupillometry showed modest ability for detecting elevated ONSD. However, its high performance in severe TBI classification warrants further evaluation. Larger, multicenter studies evaluating triage utility in prehospital and resource-limited settings are warranted to validate and extend these findings.

Для цитирования: Pant Aaradhya, Maxin Anthony, Farrokhi Farrokh, Shrestha Gentle S, Sarwal Aarti, Bohara Sandeep, Veeravagu Anand, Sharma Mohan R. Smartphone-Based Pupillometry for Noninvasive Detection of Elevated Intracranial Pressure in Nepali Acute Brain Injury Patients: A Pilot Study.. Journal of neurosurgical anesthesiology. 2026. DOI: 10.1097/ana.0000000000001117.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.