PURPOSE: We further optimized and evaluated the safety of the magnetic levator prosthesis (MLP) for temporary management of severe blepharoptosis, and compared efficacy and comfort against the ptosis crutch.

METHODS

The interpalpebral fissure (IPF) of participants (n = 12) with ptosis was measured during attempted eyelid opening, volitional closing, and spontaneous closing with no device, ptosis crutch, or the MLP. A 10-point scale documented comfort. Additionally, a 20 minute and then 1 week trial of the MLP was offered. Safety measures were skin erythema rating, change in visual acuity, and change in corneal staining.

RESULTS

The MLP and crutch opened the eye (IPF 11.2 and 9.3 mm), but the MLP allowed better volitional closure (IPF 1.0 vs. 4.9 mm, P = 0.009), but was no better in allowing spontaneous blink (IPF 7.5 vs. 7.7 mm, P = 0.722). Both devices were equally comfortable (both median 8/10 comfort, P = 0.46). With extended use, opening with the MLP showed IPF 9.24 mm at 20 minutes and 9.46 mm at 1 week, and volitional closure was IPF 0.95 and 0.52 mm, respectively. Closure on spontaneous blink improved with extended wear to IPF 5.14 and 5.18 mm, respectively (P = 0.002). Two participants exhibited moderate skin erythema and one had increased corneal staining without change in acuity.

CONCLUSIONS

The MLP is safe and feasible for temporary correction of severe ptosis.

TRANSLATIONAL RELEVANCE: First group data in patients showing successful reanimation of the eyelid with magnetic force.

Для цитирования: Houston Kevin E, Tomasi Matteo, Amaral Christina, Finch Nicole, Yoon Michael K, Lee Hang, Paschalis Eleftherios I. The Magnetic Levator Prosthesis for Temporary Management of Severe Blepharoptosis: Initial Safety and Efficacy.. Translational vision science & technology. 2018. DOI: 10.1167/tvst.7.1.7.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.