[Waldenstrom’s macroglobulinemia highlighted by turbidity interference: a case report].

Masson Pauline, Chutova Ecaterina, Oris Charlotte, Dannus Louis-Thomas, Sapin Vincent, Bouvier Damien

Клинический случайАннотация на английскомСтатьяПеревод в очереди

Аннотация

A 65-year-old female was admitted to the Emergency Department for a fall. Upon admission, a blood sample was drawn for routine laboratory tests. Blood glucose measurement using the hexokinase method was made impossible due to a high lipemia index measured by the instrument despite a clear plasma specimen. It is well known that hemolysis, icterus, and lipemia interferences disrupt spectrophotometric quantifications. However, in this case, the absence of lipemia was confirmed by plasma triglycerides measurement in the normal reference range. Protein electrophoresis along with an immunofixation were carried out to characterize a monoclonal gammopathy since the presence of a monoclonal gammopathy in very high concentrations generates turbidity and interferes with spectrophotometric assays. A gamma peak evoking a monoclonal gammopathy was found on the serum electrophoresis and the immunofixation was positive for a monoclonal IgM kappa. Myelogram showed plasmocytes, lymphoplasmacytic cells and mast cells infiltration which led to the diagnosis of Waldenstrom macroglobulinemia. Based on this case, we propose a course of action to be taken in the case of a high lipemia index and clear plasma.

Для цитирования: Masson Pauline, Chutova Ecaterina, Oris Charlotte, Dannus Louis-Thomas, Sapin Vincent, Bouvier Damien. [Waldenstrom’s macroglobulinemia highlighted by turbidity interference: a case report].. Annales de biologie clinique. 2022. DOI: 10.1684/abc.2022.1752.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.