DOI 10.1097/j.jcrs.0000000000000812
Evaluation of intraoperative aphakic eye axial length measurements using swept-source OCT.
Аннотация
PURPOSE: To evaluate intraoperative aphakic eye axial length (AL) measurements using swept-source optical coherence tomography (SS-OCT).
SETTING
Hanusch Hospital, Vienna, Austria.
DESIGN
Prospective single-center study.
METHODS
Patients scheduled for cataract surgery were measured using SS-OCT (IOLMaster 700, Carl Zeiss Meditec AG) to assess the AL. Intraoperatively (intra-OP), SS-OCT measurements were performed with a prototype device (IOLMaster 700 connected to an OPMI Lumera 700 microscope, CZM) at the beginning of cataract surgery furthermore of the aphakic eye and 2 months after surgery.
RESULTS
106 patients were included. Of the 59 eyes of 59 patients, the phakic median AL preoperatively and intra-OP was 23.61 mm ± 0.96 (standard deviation [SD]) and 23.51 mm ± 0.96 (SD), respectively. The absolute median difference was 0.028 ± 0.02 (SD) (P = .049). Median phakic AL intra-OP vs 2 months postoperatively (post-OP) was 23.51 mm ± 0.97 (SD) vs 23.49 mm ± 0.95 (SD). The absolute median difference was 0.049 ± 0.04 (SD) (P = .000). Median AL intra-OP aphakic vs 2 months post-OP pseudophakic was 23.42 mm ± 0.97 (SD) vs 23.42 mm ± 0.97 (SD), respectively. Absolute median difference was 0.038 ± 0.04 (SD) (P = .379).
CONCLUSIONS
Intra-OP, SS-OCT technology of the phakic and aphakic eye shows excellent comparability to preoperative and postoperative measurements. This technique allows AL measurements with high precision in cases in which preoperative biometric measurements are not possible.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.