OBJECTIVES

To mitigate the shortage of homograft sources, the use of handmade trileaflet expanded polytetrafluoroethylene valves in pulmonary valve replacement has shown excellent results from multicentre studies conducted in Japan. However, world-wide data outside Japan are relatively insufficient. This study presents the long-term results of a single surgeon's use of flipped-back trileaflet method in a 10-year case series.

METHODS

We have developed an efficient way to make a trileaflet-valved conduit utilizing flipped-back method for pulmonary valve replacement and have employed the technique since 2011. Retrospective data were studied between October 2010 and January 2020. Echocardiography, electrocardiogram, Pro-Brain Natriuretic Peptide and Magnetic Resonance Imaging data were analysed.

RESULTS

Fifty-five patients were reviewed and median follow-up duration was 2.9 years. The majority of diagnoses was Tetralogy of Fallot (n = 41), and these patients subsequently underwent secondary pulmonary valve replacement at a median age of 15.6 years. Survival was 92.7% with the longest follow-up period being 10 years. There was no need for reoperation, and freedom from reintervention was 98.0% at 10 years. There were 4 deaths (3 in-hospital and 1 outpatient). One patient eventually received transcatheter pulmonary valve implantation. Postoperative echocardiography showed mild or less pulmonary stenosis and pulmonary regurgitation degree in 92.2% and 92.0% of patients, respectively. Comparable magnetic resonance imaging data (n = 25) showed significant reduction in right ventricular volumes but not in ejection fractions.

CONCLUSIONS

Our series showed satisfactory long-term function of handmade flipped-back trileaflet-valved conduit used in our patients. The simple design is efficiently reproducible without complex fabrication process.

Для цитирования: Chang Te-I, Hsu Kang-Hong, Hung Wei-Li, Yeh Shu-Jen, Chen Ming-Ren, Chien Yu-San, Hsu Ron-Bin, Wang Jou-Kou, Chang Ruey-Feng, Chang Chung-I. Clinical outcomes of handmade polytetrafluoroethylene trileaflet-valved conduit used for pulmonary valve replacement.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2023. DOI: 10.1093/ejcts/ezad120.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.