DOI 10.1097/opx.0000000000000297

Медицинская библиотека · Optometry and vision science : official publication of the American Academy of Optometry · 2014

Repeatability of the modified Thorington card used to measure far heterophoria.

Cebrian Jose Luis, Antona Beatriz, Barrio Ana, Gonzalez Enrique, Gutierrez Angel, Sanchez Isabel

СтатьяПеревод в очереди

Аннотация

PURPOSE: To determine the interexaminer and intraexaminer repeatability of the modified Thorington test (TH) for distance vision in young adults and to compare these results with those observed for the heterophoria tests most commonly used in clinical practice. Agreement among tests was also assessed.

METHODS

Distance heterophoria was quantified on two separate occasions by two examiners in 110 subjects aged 18 to 32 years (mean, 19.74 years; SD, 2.5 years) using four different tests: cover test (CT) Von Graefe, Maddox rod, and modified TH. The repeatability of the tests and agreement between them was estimated by the Bland and Altman method whereby the mean difference and the 95% limits of agreement were determined as the coefficient of repeatability (COR) and coefficient of agreement.

RESULTS

The Thorington test showed best interexaminer repeatability (COR = ±1.43Δ), followed closely by CT (COR = ±1.65Δ), whereas best intraexaminer repeatability was observed for CT (COR = ±1.28Δ) followed by TH (COR = ±1.51Δ). Among the different combinations of tests, TH and CT showed best agreement indicated by the lowest coefficient of agreement (±2.23Δ) and a low mean difference (-0.63Δ) between measurements.

CONCLUSIONS

Good interexaminer and intraexaminer repeatability was observed for both TH and CT, and agreement between the two tests was also good. Given the simple administration of the TH, we recommend its clinical use to quantify distance horizontal heterophoria.

Для цитирования: Cebrian Jose Luis, Antona Beatriz, Barrio Ana, Gonzalez Enrique, Gutierrez Angel, Sanchez Isabel. Repeatability of the modified Thorington card used to measure far heterophoria.. Optometry and vision science : official publication of the American Academy of Optometry. 2014. DOI: 10.1097/opx.0000000000000297.