DOI 10.1007/s00784-026-07100-6
In vivo assessment of the Optipen optical reflectometer for detecting enamel surface changes following an erosive challenge.
Аннотация
OBJECTIVES
To evaluate whether the Optipen optical reflectometer can detect enamel surface changes associated with erosion under clinical conditions.
MATERIALS AND
METHODS
In this short-term parallel randomized clinical trial, 24 healthy participants were allocated to two groups: Test (T-rinse with a cola beverage for 30 s) and Control (C-rinse with water). Surface reflectance was measured using the Optipen optical reflectometer on the maxillary central incisors at baseline, immediately after rinsing, and after 2 h of salivary exposure, and expressed as the percentage of surface reflection intensity.
RESULTS
The test group showed significantly lower reflectance values immediately after the erosive challenge (65.4) compared with the control group (102) (p < 0.05). After 2 h of salivary exposure, no significant differences were observed between groups (T = 85.7 and C = 97.8).
CONCLUSIONS
The Optipen optical reflectometer detected immediate changes in enamel surface reflectance in vivo following a single erosive challenge.
CLINICAL RELEVANCE: This device may be useful for detecting immediate enamel surface changes following erosion. Further studies are needed to optimize reflectance measurements after salivary exposure.