MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/diagnostics16142281

Медицинская библиотека · Diagnostics (Basel, Switzerland) · 2026

Comparative Analysis of Twenty Intraocular Lens Power Calculation Formulas in Medium-Long Eyes.

Stopyra Wiktor, Voytsekhivskyy Oleksiy, Grzybowski Andrzej

СтатьяПеревод в очереди

Аннотация

Background/

Objectives

The choice of appropriate intraocular lens (IOL) power calculation formulas depends on the axial length of the eye. This study compares the accuracy of twenty formulas in medium-long eyes (24.50-25.99 mm).

Methods

Data of patients with medium-long eyes, who underwent uneventful phacoemulsification between January 2018 and September 2023, were retrospectively reviewed. Preoperative IOL power was calculated using the IOLMaster 700 with six formulas: Barrett Universal II, Haigis, Hoffer Q, Holladay 1, Holladay 2, and SRK/T. Three months postoperatively, refraction was measured. Postoperative IOL power calculations were then performed using fourteen additional formulas: Castrop, EVO 2.0, Hoffer QST, K6, Kane, Karmona, Ladas Super Formula AI (LSF AI), Naeser 2, Olsen (OLCR), Olsen (standalone), PEARL-DGS, T2, VRF CMAL, and VRF-G. The main outcome measures included standard deviation (SD) of the prediction error (PE) and the percentage of eyes with PE within ±0.25 D, ±0.50 D, ±0.75 D, and ±1.00 D.

Results

Ninety-five eyes with axial lengths ranging from 24.52 mm to 25.97 mm were included. SD values among the twenty formulas ranged from 0.179 (SRK/T) to 0.468 (Olsen OLCR). The percentage of eyes with PE within ±0.50 D ranged from 75.79% (Olsen OLCR) to 98.95% (SRK/T). The SRK/T formula, followed by Holladay 1, demonstrated significantly higher accuracy than most other formulas, while Olsen (OLCR) and Castrop were the least accurate.

Conclusions

SRK/T provided the highest accuracy in medium-long eyes, with Holladay 1 performing similarly well. All evaluated formulas achieved PE within ±0.50 D in over 74% of cases.

Для цитирования: Stopyra Wiktor, Voytsekhivskyy Oleksiy, Grzybowski Andrzej. Comparative Analysis of Twenty Intraocular Lens Power Calculation Formulas in Medium-Long Eyes.. Diagnostics (Basel, Switzerland). 2026. DOI: 10.3390/diagnostics16142281.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.