Patient-Ventilator Synchronization During Non-invasive Ventilation: A Pilot Study of an Automated Analysis System.

Letellier Christophe, Lujan Manel, Arnal Jean-Michel, Carlucci Annalisa, Chatwin Michelle, Ergan Begum, Kampelmacher Mike, Storre Jan Hendrik, Hart Nicholas, Gonzalez-Bermejo Jesus, Nava Stefano

СтатьяПеревод в очереди

Аннотация

Background

Patient-ventilator synchronization during non-invasive ventilation (NIV) can be assessed by visual inspection of flow and pressure waveforms but it remains time consuming and there is a large inter-rater variability, even among expert physicians. SyncSmart™ software developed by Breas Medical (Mölnycke, Sweden) provides an automatic detection and scoring of patient-ventilator asynchrony to help physicians in their daily clinical practice. This study was designed to assess performance of the automatic scoring by the SyncSmart software using expert clinicians as a reference in patient with chronic respiratory failure receiving NIV.

Methods

From nine patients, 20 min data sets were analyzed automatically by SyncSmart software and reviewed by nine expert physicians who were asked to score auto-triggering (AT), double-triggering (DT), and ineffective efforts (IE). The study procedure was similar to the one commonly used for validating the automatic sleep scoring technique. For each patient, the asynchrony index was computed by automatic scoring and each expert, respectively. Considering successively each expert scoring as a reference, sensitivity, specificity, positive predictive value (PPV), κ-coefficients, and agreement were calculated.

Results

The asynchrony index assessed by SynSmart was not significantly different from the one assessed by the experts (18.9 ± 17.7 vs. 12.8 ± 9.4, p = 0.19). When compared to an expert, the sensitivity and specificity provided by SyncSmart for DT, AT, and IE were significantly greater than those provided by an expert when compared to another expert.

Conclusions

SyncSmart software is able to score asynchrony events within the inter-rater variability. When the breathing frequency is not too high (<24), it therefore provides a reliable assessment of patient-ventilator asynchrony; AT is over detected otherwise.

Для цитирования: Letellier Christophe, Lujan Manel, Arnal Jean-Michel, Carlucci Annalisa, Chatwin Michelle, Ergan Begum, Kampelmacher Mike, Storre Jan Hendrik, Hart Nicholas, Gonzalez-Bermejo Jesus, Nava Stefano. Patient-Ventilator Synchronization During Non-invasive Ventilation: A Pilot Study of an Automated Analysis System.. Frontiers in medical technology. 2021. DOI: 10.3389/fmedt.2021.690442.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.