MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.athoracsur.2023.08.019

Медицинская библиотека · The Annals of thoracic surgery · 2023

Long-term Outcomes of Polytetrafluoroethylene Bicuspid Pulmonary Valve Replacement.

Kim Eung Re, Lee Cheul, Lee Chang-Ha, Cho Sang-Hoon

СтатьяПеревод в очереди

Аннотация

BACKGROUND

In 2016 we reported promising midterm outcomes of bicuspid pulmonary valve replacement using 0.1-mm polytetrafluoroethylene (PTFE) membrane. This follow-up study analyzes long-term outcomes and risk factors for reintervention and structural valve deterioration (SVD).

METHODS

We performed a retrospective review of the original 119 patients who underwent PTFE bicuspid pulmonary valve replacement. Median patient age was 16.9 years (range, 0.4-57.1). Reintervention was defined as any surgical or percutaneous catheter procedure on the PTFE valve. SVD was defined as development of a peak pressure gradient ≥ 50 mm Hg or at least a moderate amount of pulmonary regurgitation on follow-up echocardiography.

RESULTS

The median follow-up duration was 9.5 years. The survival rate was 96.5% at 5 and 10 years, with 2 early and 2 late mortalities. Freedom from reintervention was 90.0% at 5 years and 63.3% at 10 years. Freedom from SVD was 92.8% at 5 years and 51.1% at 10 years, with regurgitation the predominant mode (64.6%). Freedom from both reintervention and SVD at 5 and 10 years were 89.1% and 49.5%, respectively. Multivariable analysis identified smaller valve diameter (hazard ratio, 0.82; P < .001) and more than trivial pulmonary regurgitation at discharge (hazard ratio, 5.81; P < .001) as risk factors for reintervention or SVD.

CONCLUSIONS

Long-term results of the PTFE bicuspid pulmonary valve replacement were acceptable. However, improvements may be needed to reduce technical error and improve durability. Smaller valve diameter and more than trivial pulmonary regurgitation at discharge were risk factors for reintervention or SVD, warranting careful follow-up for timely reintervention.

Для цитирования: Kim Eung Re, Lee Cheul, Lee Chang-Ha, Cho Sang-Hoon. Long-term Outcomes of Polytetrafluoroethylene Bicuspid Pulmonary Valve Replacement.. The Annals of thoracic surgery. 2023. DOI: 10.1016/j.athoracsur.2023.08.019.