MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1161/circinterventions.120.009707

Медицинская библиотека · Circulation. Cardiovascular interventions · 2021

Mid-Term Outcomes Following Percutaneous Pulmonary Valve Implantation Using the "Folded Melody Valve" Technique.

Jalal Zakaria, Valdeolmillos Estíbaliz, Malekzadeh-Milani Sophie, Eicken Andreas, Georgiev Stanimir, Hofbeck Michael, Sieverding Ludger, Gewillig Marc, Ovaert Caroline, Bouvaist Helene, Pillois Xavier, Thambo Jean-Benoit, Boudjemline Younes

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The folded valve is a manual shortening of the Melody device, which has been validated as a valuable therapeutic option for the management of dysfunctional right ventricular outflow tracts needing a short valved stent. In this article, we aimed to evaluate, in a multicenter cohort, the mid-term outcomes of patients in whom a percutaneous pulmonary valve implantation was performed using the folded valve technique.

METHODS

A 2012 to 2018 retrospective multicenter study was performed in 7 European institutions. All patients who benefit from percutaneous pulmonary valve implantation with a folded Melody valve were included.

RESULTS

A total of 49 patients (median age, 19 years [range 4–56], 63% male) were included. The primary percutaneous pulmonary valve implantation indication was right ventricular outflow tract stenosis (n=19; 39%), patched native right ventricular outflow tracts were the most common substrate (n=15; 31%). The folded technique was mostly used in short right ventricular outflow tracts (n=28; 57%). Procedural success was 100%. After a median follow-up of 28 months (range, 4–80), folded Melody valve function was comparable to the immediate postimplantation period (mean transvalvular peak velocity=2.6±0.6 versus 2.4±0.6 m/s, P>0.1; only 2 patients had mild pulmonary regurgitation). Incidence rate of valve-related reinterventions was 2.1% per person per year (95% CI, 0.1%–3.9%). The probability of survival without valve-related reinterventions at 36 months was 90% (95% CI, 76%–100%).

CONCLUSIONS

The folded Melody valve is a safe technique with favorable mid-term outcomes up to 6.5 years after implantation, comparable with the usual Melody valve implantation procedure. Complications and reinterventions rates were low, making this technique relevant in selected patients.

Для цитирования: Jalal Zakaria, Valdeolmillos Estíbaliz, Malekzadeh-Milani Sophie, Eicken Andreas, Georgiev Stanimir, Hofbeck Michael, Sieverding Ludger, Gewillig Marc, Ovaert Caroline, Bouvaist Helene, Pillois Xavier, Thambo Jean-Benoit, Boudjemline Younes. Mid-Term Outcomes Following Percutaneous Pulmonary Valve Implantation Using the "Folded Melody Valve" Technique.. Circulation. Cardiovascular interventions. 2021. DOI: 10.1161/circinterventions.120.009707.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.