m+Medelevator.RU
← Поиск публикаций
American journal of veterinary research · 2022

Transpulmonary ultrasound dilution is an acceptable technique for cardiac output measurement in anesthetized pigs.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Paranjape Vaidehi V, Shih Andre C, Garcia-Pereira Fernando L, Saksena Siddharth

Аннотация

OBJECTIVE: To evaluate cardiac output (CO) measurements using transpulmonary ultrasound (TPUD) technology and compare results with those of the gold standard, pulmonary arterial catheter thermodilution (PACTD), in 6 healthy anesthetized pigs during acute hemodynamic changes caused by manipulation of the blood volume.

ANIMALS: 6 healthy male Landrace pigs.

PROCEDURES: Over a period of 1 week, pigs were anesthetized with isoflurane, mechanically ventilated, and underwent instrumentation in dorsal recumbency. They were subjected to sequential experimental states during which the blood volume was manipulated so that the animals transitioned from normovolemia to hypovolemia (20% and 40% of blood volume depletion), back to normovolemia (autologous blood transfusion), and then to hypervolemia (following colloid bolus). During each volume state, CO measurements were compared between TPUD and PACTD.

RESULTS: The mean ± SD relative bias between TPUD and PACTD was 7.71% ± 21.2% with limits of agreement -33.9% to 49.3%, indicating TPUD slightly underestimated CO values, compared with values obtained with PACTD. The mean ± SD of the bias between the 2 methods was 0.13 ± 0.5 L/min. Only 5 of 36 (13.9%) TPUD CO measurements had an absolute value of relative bias > 30%. The percentage error calculated for TPUD was 29.4%.

CLINICAL RELEVANCE: Results suggested that TPUD measurements have acceptable agreement with PACTD measurements. Moreover, TPUD exhibits promising potential in being used interchangeably with PACTD for future hemodynamic research involving swine as species of interest.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049