MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/j.jcrs.0000000000000944

Медицинская библиотека · Journal of cataract and refractive surgery · 2022

Correlation between Placido-disc and rotating Scheimpflug keratometric findings before and after corneal crosslinking in children with keratoconus.

Polido Júlia, Xavier Dos Santos Araújo Maria Emília, H Wakamatsu Tais, T Lopes Bernardo, Alexander João G, Cabral Thiago, Ambrósio Renato, Freitas Denise

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate correlation between Placido-disc and rotating Scheimpflug keratometric findings in children with progressive keratoconus (KC) before and after corneal crosslinking (CXL) and investigate whether these limits of agreement varied according to disease severity.

SETTING

Department of Ophthalmology of São Paulo Hospital, São Paulo, Brazil.

DESIGN

Prospective nonrandomized open study.

METHODS

Data obtained using rotating Scheimpflug-based tomography and Placido-disc-based topography devices were collected from preoperative and last follow-up postoperative children with KC operated on using standard CXL protocol. Correlation and agreement analyses were performed between the 2 devices before and after CXL to obtain keratometric (K) findings.

RESULTS

44 eyes from 44 patients aged 8 to 16 years were analyzed at all timepoints. All parameters were found to be strongly correlated before ( r = 0.84 to 0.99, P < .001) and after ( r = 0.93 to 0.99, P < .001) CXL. The mean Scheimpflug measurements of flat K, steep K, Kmax, mean K, and corneal astigmatism were higher than Placido-disc measurements in a preoperative period. This mean difference decreased in postoperative, but, with exception of Kmax and corneal astigmatism, Scheimpflug measurements remained higher. The mean parameter measurements from both devices decreased after CXL; 95% limits of agreement between instruments were wide for all parameters and decreased in postoperative and in mild KC.

CONCLUSIONS

Keratometry measurements obtained using rotating Scheimpflug and Placido-disc technology were found to be closely correlated but not interchangeable before and after CXL in pediatric patients. Agreement between devices was better after CXL and in mild KC than in advanced KC.

Для цитирования: Polido Júlia, Xavier Dos Santos Araújo Maria Emília, H Wakamatsu Tais, T Lopes Bernardo, Alexander João G, Cabral Thiago, Ambrósio Renato, Freitas Denise. Correlation between Placido-disc and rotating Scheimpflug keratometric findings before and after corneal crosslinking in children with keratoconus.. Journal of cataract and refractive surgery. 2022. DOI: 10.1097/j.jcrs.0000000000000944.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.