DOI 10.1161/01.cir.98.20.2160

Медицинская библиотека · Circulation · 1998

Influence of lead selection and population on automated measurement of QT dispersion.

Macfarlane P W, McLaughlin S C, Rodger J C

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The study of QT dispersion (QTd) is of increasing clinical interest, but there are very few data in large healthy populations. Furthermore, there is still discussion on the extent to which QTd reflects dispersion of measurement. This study addresses these problems.

METHODS AND

RESULTS

Twelve-lead ECGs recorded on 1501 apparently healthy adults and 1784 healthy neonates, infants, and children were used to derive normal limits of QTd and QT intervals by use of a fully automated approach. No age gradient or sex differences in QTd were seen and it was found that an upper limit of 50 ms was highly specific. Three-orthogonal-lead ECGs (n=1220) from the Common Standards for Quantitative Electrocardiography database were used to generate derived 12-lead ECGs, which had a significant increase in QTd of 10.1+/-13.1 ms compared with the original orthogonal-lead ECG but a mean difference of only 1.63+/-12.2 ms compared with the original 12-lead ECGs. In a population of 361 patients with old myocardial infarction, there was a statistically significant increase in mean QTd compared with that of the adult normal group (32.7+/-10.0 versus 24.53+/-8.2 ms; P<0. 0001). An estimate of computer measurement error was also obtained by creating 2 sets of 1220 ECGs from the original set of 1220. The mean error (difference in QTd on a paired basis) was found to be 0. 28+/-9.7 ms.

CONCLUSIONS

These data indicate that QTd is age and sex independent, has a highly specific upper normal limit of 50 ms, is significantly lower in the 3-orthogonal-lead than in the 12-lead ECG, and is longer in patients with a previous myocardial infarction than in normal subjects.

Для цитирования: Macfarlane P W, McLaughlin S C, Rodger J C. Influence of lead selection and population on automated measurement of QT dispersion.. Circulation. 1998. DOI: 10.1161/01.cir.98.20.2160.