Assessment of Contrast-Enhanced Ultrasound Detection Limit for Abdominal Hemorrhage.

Gupta Jay F, Candel Ion, Schoen Scott, Ottensmeyer Mark P, Arshad Saaid H, Iqbal Shams I, Wang Xiaohong, Naja Kim, Waugh Mary, Pierce Theodore T, Telfer Brian A, Samir Anthony E

СтатьяПеревод в очереди

Аннотация

Hemorrhage is the leading cause of preventable death from trauma, with most fatalities occurring in prehospital settings. The current standard for point-of-care hemorrhage detection is the focused assessment with sonography in trauma (FAST) exam, or computed tomography (CT) angiography after hospital arrival. Contrast-enhanced ultrasound (CEUS) has been proposed as a next-generation FAST exam, due to potential for: 1) differentiating active versus inactive bleeding; 2) localizing bleeding sources; and 3) guiding percutaneous and/or endovascular treatments to temporize bleeding for prolonged transport. While these advantages have been reported in human trauma case studies, characterization of the minimum hemorrhage rate detectable by CEUS remains a translational research need. Prior work using conventional B-mode imaging and microvascular CEUS algorithms showed the detection of 5 mL/min flow in a generic hemorrhage phantom. Here, we apply nonlinear imaging waveforms [pulse inversion (PI) and amplitude modulation (AM)] with an iterative spatiotemporal filter that further enhances contrast signals [fast iterative shrinkage-thresholding algorithm (FISTA)] to detect significantly lower flow rates in an anatomically realistic abdominal hemorrhage phantom. We demonstrate detection of bleeding at a minimum flow rate of 0.5 mL/min, comparable to multidetector CT. We achieve this detection limit even with 10% of the standard contrast dose (area under the receiver operating characteristic (ROC) curve of 0.90 and 0.84 for PI + FISTA and AM + FISTA, respectively, compared to 0.38 for conventional B-mode), suggesting that CEUS examination is feasible throughout contrast wash-in/wash-out. Finally, to bridge toward in vivo characterization, we present a novel porcine liver hemorrhage model tailored to evaluating CEUS capabilities.

Для цитирования: Gupta Jay F, Candel Ion, Schoen Scott, Ottensmeyer Mark P, Arshad Saaid H, Iqbal Shams I, Wang Xiaohong, Naja Kim, Waugh Mary, Pierce Theodore T, Telfer Brian A, Samir Anthony E. Assessment of Contrast-Enhanced Ultrasound Detection Limit for Abdominal Hemorrhage.. IEEE transactions on ultrasonics. 2026. DOI: 10.1109/tuson.2026.3677094.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.