Effects of a modified passive socket system on short-term changes in residuum volume and comfort: A preliminary study in transtibial amputees.

Atallah Huthaifa, Kenney Laurence, Howard David, Liu Anmin, Head John

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Changes in residuum volume are a common problem in lower-limb amputees during prosthesis usage, and can lead to poor suspension, impaired gait, and tissue damage. Residuum volume can be affected by the in-socket air pressure, which will influence fluid flow in and around the residuum. The use of "active" pumps to reduce air pressure has been shown to conserve the residuum volume, but these are expensive and unlikely to be widely available. An alternative, passive approach, based on Boyles' law, is to introduce a larger distal void volume at the end of the socket and hence reduce the change in pressure for a given change in volume.

OBJECTIVES

To compare the performance across three test conditions (passive-conventional, with standard distal void; passive-with increased distal void; and active system) in terms of residuum volume changes and comfort.

STUDY

DESIGN

Repeated-measures experiment under three test conditions.

METHODS

Five transtibial amputee participants (three males and two females), aged between 27 and 67 years, and of mobility grade K2 or K3, were fitted with a bespoke test prosthesis that was adapted to include the three test conditions. Residuum volume was measured before and after walking under each test condition (presented in a random order). Comfort was also assessed after walking with each test condition.

RESULTS

The reduction in residuum volume, relative to the baseline volume, was higher for the conventional passive system (4.2% ± 2.8%) compared with the modified passive (1.4% ± 1.4%) and active (1.6% ± 1.1%) systems.

CONCLUSION

The use of a passive suspension system with an increased distal void within the socket may help to stabilize the residuum volume during prosthesis usage.

Для цитирования: Atallah Huthaifa, Kenney Laurence, Howard David, Liu Anmin, Head John. Effects of a modified passive socket system on short-term changes in residuum volume and comfort: A preliminary study in transtibial amputees.. Prosthetics and orthotics international. 2022. DOI: 10.1097/pxr.0000000000000053.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.