Influence of guided waves in bone on pulse-inversion contrast-enhanced ultrasound.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wang Diya, Zhang Xinyu, Sang Yuchao, Qu Zhen, Su Qiang, Zhao Jing, Wan Mingxi
Аннотация
PURPOSE: Guided waves generated from bone cortex inevitably act on microbubbles flowing through skeletal muscle capillaries in contrast-enhanced ultrasound (CEUS) and might influence the image quality. However, the action mechanism underlying the guided waves influence is still unknown, especially under contrast pulse-inversion transmission mode. This study aimed to clarify the influence of guided waves on pulse-inversion CEUS, which was investigated via in vitro infusion experiments.
METHOD: Tibia guided waves were detected at pulse-inversion transmission and then characterized by using a short-time Fourier transform energy distribution. Using results at normal incidence as a baseline, the influence of guided wave dispersion on the contrast and resolution of pulse-inversion CEUS was investigated at an oblique incidence through continuous microbubbles infusion experiments in a vessel-tibia flow phantom.
RESULTS: Frequency-dispersive property of tibia guided waves was observed at phases 0° and 180°, which improved the contrast of CEUS and reduced its resolution. Pulse-inversion CEUS balanced the contrast enhancement and resolution degeneration induced by guided waves. By contrast, contrast-to-tissue ratio of pulse-inversion CEUS increased by up to 109.1 ± 13.2% (P < 0.05) due to guided waves and its resolution was up to 0.9 ± 0.1 times that of baseline.
CONCLUSIONS: Alterations of contrast and resolution in pulse-inversion CEUS induced by guided waves might provide an additional assessment for the capillary perfusion in the skeletal muscle near the bone cortex.
METHOD: Tibia guided waves were detected at pulse-inversion transmission and then characterized by using a short-time Fourier transform energy distribution. Using results at normal incidence as a baseline, the influence of guided wave dispersion on the contrast and resolution of pulse-inversion CEUS was investigated at an oblique incidence through continuous microbubbles infusion experiments in a vessel-tibia flow phantom.
RESULTS: Frequency-dispersive property of tibia guided waves was observed at phases 0° and 180°, which improved the contrast of CEUS and reduced its resolution. Pulse-inversion CEUS balanced the contrast enhancement and resolution degeneration induced by guided waves. By contrast, contrast-to-tissue ratio of pulse-inversion CEUS increased by up to 109.1 ± 13.2% (P < 0.05) due to guided waves and its resolution was up to 0.9 ± 0.1 times that of baseline.
CONCLUSIONS: Alterations of contrast and resolution in pulse-inversion CEUS induced by guided waves might provide an additional assessment for the capillary perfusion in the skeletal muscle near the bone cortex.