Significant dose reduction employing dual-low-dose PET/CT in pediatric lymphoma: maintaining diagnostic, staging, and response-assessing efficacy.

Huang Yuqian, Tang Si, Wang Tianquan, Li Shatong, Cao Liuyi, Chen Wanqi, Liu Lei, Wu Tingfan, Zhao Yumo, Fan Wei, Li Yinghe, Zhang Yizhuo, Hu Yingying

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To evaluate the clinical efficacy of a dual-low-dose total-body PET/CT protocol in pediatric lymphoma patients, incorporating low-dose PET imaging in combination with two distinct low-dose CT acquisition strategies.

MATERIALS AND

METHODS

Pediatric lymphoma patients (≤ 18 years) receiving half-dose [18F]FDG (1.85 ± 0.2 MBq/kg) were prospectively enrolled and divided into low-dose CT group (LDCT) and ultra-low-dose CT group (ULDCT) using uEXPLORER scanner. Image quality was evaluated based on subjective and objective metrics. The reference standard of lesions was confirmed with pathological and follow-up evidence. The effective dose (ED), the diagnostic efficacy for detecting lesions, and the accuracy for staging, treatment responses were compared. Institutional Review Board approval was obtained.

RESULTS

A total of 246 pediatric patients were enrolled (9.2 ± 3.8 years, 180 male). The LDCT (n = 127) and ULDCT (n = 119) had similar clinical characteristics (p > 0.05) except for male percentage (78.7% vs. 67.2%, p = 0.042). The total ED of ULDCT was significantly lower than that of LDCT (5.6 ± 1.9 mSv vs. 9.7 ± 3.2 mSv, p < 0.001). Both groups showed acceptable subject image quality and were comparable in main objective parameters. ULDCT showed a sensitivity lower than LDCT (99.0% vs. 99.9%, p = 0.010) in diagnosing lymph node lesions, while their specificity (96.0% vs. 94.3%) and area under curve (AUC, 0.978 vs. 0.982) were similar (both p > 0.05). LDCT and ULDCT were also comparable in diagnosing extranodal involvements (AUCs > 0.800), staging accuracy (94.6% and 93.6%), and response assessment accuracy (97.6% and 96.3%).

CONCLUSIONS

Dual-low-dose total-body PET/CT allowed a significant reduction in total ED without compromising imaging quality and diagnosis performance in pediatric lymphoma patients.

KEY POINTS: Question [18F]FDG PET/CT scans are crucial for pediatric lymphoma patients; however, an optimized low-dose strategy has not been achieved. Findings Dual-low-dose PET/CT demonstrated sufficient image quality and clinical efficacy in pediatric lymphoma patients, with significantly reduced radiation exposure. The total radiation dosage of the ultra-low-dose CT protocol was 5.6 ± 1.9 mSv. Clinical relevance A dual-low-dose PET/CT protocol is an optimized imaging strategy for pediatric lymphoma patients that significantly reduces radiation exposure while providing sufficient clinical efficacy for diagnosing, staging and treatment response evaluation.

Для цитирования: Huang Yuqian, Tang Si, Wang Tianquan, Li Shatong, Cao Liuyi, Chen Wanqi, Liu Lei, Wu Tingfan, Zhao Yumo, Fan Wei, Li Yinghe, Zhang Yizhuo, Hu Yingying. Significant dose reduction employing dual-low-dose PET/CT in pediatric lymphoma: maintaining diagnostic, staging, and response-assessing efficacy.. European radiology. 2025. DOI: 10.1007/s00330-025-11847-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.