Technical strategies for radiation dose optimisation in paediatric pelvic radiography: a systematic review.

Kamath Koustubh, Sukumar Suresh, David Leena R, Kadavigere Rajagopal, Shah Hitesh Hasmukhlal, Panakkal Nitika C, Pradhan Abhimanyu, Dkhar Winniecia, Planiswamy Hari Prakash, Ravichandran Sneha, Shettigar Dilip, Muthu Sathya Sabina

СтатьяСистематический обзорПеревод в очереди

Аннотация

Paediatric pelvic radiography is a frequently performed diagnostic procedure that exposes radiosensitive organs to ionising radiation. Optimisation of imaging parameters is essential to minimise radiation dose while preserving diagnostic image quality. This systematic review aimed to evaluate technical strategies for radiation dose optimisation in paediatric pelvic radiography and assess their impact on radiation dose and image quality. A systematic search of PubMed, Scopus, Embase, and Web of Science was conducted for studies published between 2004 and 2025. Experimental and clinical studies assessing radiation dose and image quality in paediatric pelvic radiography were included. Data on exposure parameters, dose metrics, and image quality assessment methods were extracted and synthesised narratively. Nine studies met the inclusion criteria, predominantly comprising phantom-based experimental designs. Increasing tube potential (kVp), extending source-to-image distance (SID), and applying additional aluminium or copper filtration consistently reduced radiation dose. Dose reductions of up to 60-70% were reported without clinically significant loss of image quality. Although objective image quality metrics (e.g., CNR, SNR) occasionally declined, subjective assessments confirmed that images remained diagnostically acceptable. Optimisation of technical parameters in paediatric pelvic radiography can achieve substantial radiation dose reduction while maintaining clinically acceptable image quality. However, the predominance of phantom-based evidence highlights the need for further patient-based clinical studies to support routine implementation.

Для цитирования: Kamath Koustubh, Sukumar Suresh, David Leena R, Kadavigere Rajagopal, Shah Hitesh Hasmukhlal, Panakkal Nitika C, Pradhan Abhimanyu, Dkhar Winniecia, Planiswamy Hari Prakash, Ravichandran Sneha, Shettigar Dilip, Muthu Sathya Sabina. Technical strategies for radiation dose optimisation in paediatric pelvic radiography: a systematic review.. Radiological physics and technology. 2026. DOI: 10.1007/s12194-026-01055-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.