Precision of bone density and micro-architectural properties at the distal radius and tibia in children: an HR-pQCT study.

Kawalilak C E, Bunyamin A T, Björkman K M, Johnston J D, Kontulainen S A

СтатьяПеревод в очереди

Аннотация

UNLABELLED: Precision errors need to be known when monitoring bone micro-architecture in children with HR-pQCT. Precision errors for trabecular bone micro-architecture ranged from 1 to 8% when using the standard evaluation at the radius and tibia. Precision errors for cortical bone micro-architecture ranged from 1 to 11% when using the advanced cortical evaluation.

INTRODUCTION

Our objective was to define HR-pQCT precision errors (CV%RMS) and least significant changes (LSCs) at the distal radius and tibia in children using the standard evaluation and the advanced cortical evaluation.

METHODS

We scanned the distal radius (7% of ulnar length) and tibia (8% of tibia length) of 32 children (age range 8-13; mean age 11.3; SD 1.6 years) twice (1 week apart) using HR-pQCT (XtremeCT1). We calculated root-mean-squared coefficients of variation (CV%RMS) to define precision errors and LSC to identify differences required to detect change.

RESULTS

Precision errors ranged between 1-8 and 1-5% for trabecular bone outcomes (obtained with standard evaluation) and between 1.5-11 and 0.5-6% for cortical bone outcomes (obtained with advanced cortical evaluation) at the distal radius and tibia, respectively. Related LSCs ranged between 3-21 and 3-14% for trabecular bone outcomes and between 4-30 and 2-16% for cortical bone outcomes at the distal radius and tibia, respectively.

CONCLUSIONS

HR-pQCT precision errors were between 1 and 8% (LSC 3-21%) for trabecular bone outcomes and 1 and 11% (LSC 2-30%) for cortical bone outcomes at the radius and tibia in children. Cortical bone outcomes obtained using the advanced cortical evaluation appeared to have lower precision errors than cortical outcomes derived using the standard evaluation. These findings, combined with better-defined cortical bone contours with advanced cortical evaluation, indicate that metrics from advanced cortical evaluation should be utilized when monitoring cortical bone properties in children.

Для цитирования: Kawalilak C E, Bunyamin A T, Björkman K M, Johnston J D, Kontulainen S A. Precision of bone density and micro-architectural properties at the distal radius and tibia in children: an HR-pQCT study.. Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA. 2017. DOI: 10.1007/s00198-017-4185-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.