Медицинская библиотека · Journal of X-ray science and technology · 2022

Dosimetric comparison of volumetric modulated arc therapy with TrueBeam LINAC and hypofractionated radiosurgery with gamma knife ICON for large (>10 cm3) skull base meningiomas.

Saglam Yucel, Samanci Yavuz, Bolukbasi Yasemin, Peker Selcuk

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Multi Fractionated stereotactic radiosurgery (MF-SRS) of Linac has an essential role in the treatment of skull base meningiomas (sbMNG). However, Gamma Knife Icon (GK) allows MF-SRS using mask immobilization with onboard image guidance.

OBJECTIVE

This dosimetric study aims to investigate whether equivalent plan quality can be achieved with Volumetric Modulated Arc Therapy (VMAT) in patients with large sbMNG (>10 cm3) previously treated with GK.

METHODS

Twenty patients with the median target volume of 19.7cm3 are re-planned by using VMAT with 20 Gy in 5 fractions. Plan qualities are compared to tumor coverage, paddick conformity index (PCI), gradient index (GI), V4 Gy, V10 Gy, V12 Gy, optic chiasm V20 Gy, brainstem V23 Gy, optic nerve V25 Gy volumes, and maximum doses for all. Additionally, beam-on time and approximate planning time are also analyzed and compared.

RESULTS

All plans provide adequate clinical requirements. First, the CI is comparable for the GK and VMAT (0.99±0.01 vs. 1.13±0.20; p = 0.18). Second, VMAT has a significantly higher GI than GK (3.81±0.35 vs. 2.63±0.09; p < 0.001). Third, the PCI is significantly higher in GK than VMAT (0.76±0.05 vs. 0.70±0.07; p < 0.001). The lower GI of the GK also results in significantly lower V4 Gy (156.1±43.8 vs. 207.5±40.1 cm3, p < 0.001) and V10 Gy (26.1±9.0 vs. 28.9±7.7 cm3, p < 0.001) compared to VMAT. Last, the VMAT reduces beam-on time (4.8±0.5 vs. 19±1.1 min.; p < 0.001).

CONCLUSION

Although both systems have succeeded in creating effective plans in clinical practice, the GK reveals more effective lower normal brain tissue doses. However, the shorter treatment time with LINAC, excluding the total procedure time, can be considered advantageous over GK.

Для цитирования: Saglam Yucel, Samanci Yavuz, Bolukbasi Yasemin, Peker Selcuk. Dosimetric comparison of volumetric modulated arc therapy with TrueBeam LINAC and hypofractionated radiosurgery with gamma knife ICON for large (>10 cm3) skull base meningiomas.. Journal of X-ray science and technology. 2022. DOI: 10.3233/xst-221264.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.