Biomechanical performance of static and NiTiNOL compression devices in first metatarsophalangeal joint arthrodesis.

Johnson James, King Shannon, Gregory Kristen, Kline Courtney, Gonzalez Tyler, Jackson J Benjamin, Safranski David L

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

First metatarsophalangeal (MTP) joint arthrodesis is commonly performed to treat hallux rigidus. Traditional stainless steel or titanium constructs (e.g., locking plates with screws) provide sufficient stiffness but are prone to permanent deformation after excessive pre-fusion loading, leading to plantar gapping and malalignment. Superelastic NiTiNOL staples capable of sustained dynamic compression (SDC) have been developed to address these shortcomings.

PURPOSE: To compare the biomechanical performance of first MTP arthrodesis constructs using static fixation, SDC staples, or hybrid static/SDC fixation.

STUDY

DESIGN

Controlled laboratory study using synthetic MTP joint replicas.

METHODS

Solid foam first MTP models (N=6/group) underwent arthrodesis with one of five constructs. Specimens were cyclically loaded for 1,000 cycles at physiologic loads, followed by destructive displacement testing. Bending stiffness, plantar gapping at simulated walking loads, and permanent plantar gapping were quantified. One-way ANOVA comparisons were performed with the locking plate and screw construct as the control.

RESULTS

Bending stiffness and plantar gapping at simulated walking loads were not significantly different across groups (p>.07). However, all SDC constructs demonstrated significantly reduced permanent plantar gapping compared with locking plate and screw (≤0.29 mm vs. 0.85 mm, p<.001). The four-leg Y staple alone or in combination with a static screw provided the highest bending stiffness and greatest reduction in permanent plantar gapping.

CONCLUSION

Superelastic NiTiNOL SDC staples yield equivalent construct stiffness and markedly reduce permanent plantar gapping compared with static fixation. Hybrid constructs may further enhance biomechanical stability, supporting future clinical evaluation of SDC devices in first MTP arthrodesis.

Для цитирования: Johnson James, King Shannon, Gregory Kristen, Kline Courtney, Gonzalez Tyler, Jackson J Benjamin, Safranski David L. Biomechanical performance of static and NiTiNOL compression devices in first metatarsophalangeal joint arthrodesis.. The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons. 2025. DOI: 10.1053/j.jfas.2025.12.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.