MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s43390-025-01134-8

Медицинская библиотека · Spine deformity · 2025

Rod options to outcomes: comparing sagittal correction in pediatric posterior spinal fusion by rod size and material.

Todderud Julia E, Nugraha Hans K, Kelly Michael P, Pahys Joshua, Shah Suken, Fletcher Nicholas, Newton Peter O, Sucato Daniel, Yaszay Burt, Larson A Noelle, Harms Study Group

СтатьяМногоцентровое исследованиеСравнительное исследованиеПеревод в очереди

Аннотация

PURPOSE: Our study aims to evaluate the impact of rod diameter and material on sagittal plane correction and reoperation. We hypothesize larger diameter rods would improve the sagittal plane alignment without increasing complications or progression of proximal junctional kyphosis (PJK).

METHODS

Data were reviewed from consecutive series of Lenke 1-4 AIS patients enrolled in a multicenter registry between 2010 and 2019 with minimum 2-year follow-up. Patients who underwent PSF with 5 common rod types (5.5 or 6.0 cobalt chrome, 5.5 stainless steel, 5.5 or 6.0 titanium rods) were evaluated. Other rod types/diameters were excluded from this study.

RESULTS

1348 patients treated by 29 surgeons at 13 centers met inclusion criteria. 42 had 5.5 titanium rods (3%), 651 patients had 5.5 cobalt chrome (48%), 586 had 5.5 stainless steel (43%), 46 had 6.0 titanium (3%), and 23 had 6.0 cobalt chrome (2%). Preoperatively there was no difference in curve magnitude or flexibility. Patients that received stiffer rods were older and taller. At 2 years, the 5.5 stainless steel patients had the greatest major curve correction (62.5%) while 5.5 titanium patients had the lowest correction (54.2%) (whole cohort p < 0.001). At 2-year follow-up there was most improved T2-T12 kyphosis in the 6.0 titanium group (+ 6.5°), and least in the 5.5 titanium group (-3.2°) (p = 0.014). T5-T12 changes were greatest in the 6.0 titanium group (+ 3.9°) and lowest in the 5.5 titanium group (-6.7°) (p < 0.001). PJK increased most in the 5.5 titanium cohort (+ 4.0) and least in the 6.0 cobalt chrome cohort (+ 0.8) (p < 0.001). Complication rates were highest in the 5.5 stainless steel patients (12.6%) while rates of reoperations were highest in the 5.5 cobalt chrome cohort (n = 9, 1.4%).

CONCLUSION

In this retrospective review of patients undergoing PSF with rods of varying size and materials, there was evidence of better restoration of T5-T12 kyphosis with stiffer/larger rods without increased risk of PJK progression.

Для цитирования: Todderud Julia E, Nugraha Hans K, Kelly Michael P, Pahys Joshua, Shah Suken, Fletcher Nicholas, Newton Peter O, Sucato Daniel, Yaszay Burt, Larson A Noelle, Harms Study Group. Rod options to outcomes: comparing sagittal correction in pediatric posterior spinal fusion by rod size and material.. Spine deformity. 2025. DOI: 10.1007/s43390-025-01134-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.