MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00586-025-08882-1

Медицинская библиотека · European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2025

Factors causing deviation of robot-assisted pedicle screw placement for adolescent idiopathic scoliosis: why does deviation still occur even with robotics?

Ueno Jun, Torii Yoshiaki, Iinuma Masahiro, Yoshida Atsuhiro, Tomochika Ken, Hideshima Takahiro, Ito Makoto, Akazawa Tsutomu

СтатьяПеревод в очереди

Аннотация

PURPOSE: To clarify the factors that are likely to cause deviation in robot-assisted pedicle screw placement in surgery for adolescent idiopathic scoliosis (AIS).

METHODS

Thirty-two consecutive patients with AIS who underwent posterior spinal fusion using a spine robotics system were included. Robot-assisted pedicle screw placements were evaluated using Gertzbein-Robbins system. Deviation rates for breaches of 2 mm or more (Grade C, D, and E) were calculated. Pedicle channel grade, vertebral level, distance from the reference frame (RF), insertion order of the screws, curve laterality, and use of a navigated high-speed drill were investigated as factors causing deviation.

RESULTS

A total of 465 robot-assisted pedicle screws were placed, and deviations were observed in 13 (2.8%). The deviation rates were significantly different by pedicle channel grade (Grade 1: 0%, Grade 2: 0%, Grade 3: 3.1%, Grade 4: 20.4%, p < 0.001), vertebral level, and use of a high-speed drill. There were no significant differences in deviation rates for distance from the RF, insertion order of the screws, or curve laterality. Logistic regression analysis showed pedicle channel grade 4 was a significant risk factor for deviation.

CONCLUSION

The factor that was most likely to cause deviation in robot-assisted pedicle screw placement in AIS was pedicle channel grade. The most likely situation to cause deviation was a narrow pedicle with an internal diameter of less than one mm. Surgeons should be aware that even with the use of robotics, placing screws into such narrow pedicles is challenging.

Для цитирования: Ueno Jun, Torii Yoshiaki, Iinuma Masahiro, Yoshida Atsuhiro, Tomochika Ken, Hideshima Takahiro, Ito Makoto, Akazawa Tsutomu. Factors causing deviation of robot-assisted pedicle screw placement for adolescent idiopathic scoliosis: why does deviation still occur even with robotics?. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2025. DOI: 10.1007/s00586-025-08882-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.